DateProductReasonClassFirm
Jul 29, 2025 Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contai... VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencin... Class II Ortho-Clinical Diagnostics, Inc.
Jul 29, 2025 VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit =... VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencin... Class II Ortho-Clinical Diagnostics, Inc.
Jul 9, 2025 VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261; Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Conditio... Class II Ortho-Clinical Diagnostics, Inc.
Jun 12, 2025 VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITR... Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than exp... Class II Ortho-Clinical Diagnostics, Inc.
Jun 10, 2025 VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 ... Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain... Class II Ortho-Clinical Diagnostics, Inc.
Jun 10, 2025 VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 sli... Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain... Class II Ortho-Clinical Diagnostics, Inc.
Jun 10, 2025 Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Product... Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain... Class II Ortho-Clinical Diagnostics, Inc.
Jun 10, 2025 Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Product... Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain... Class II Ortho-Clinical Diagnostics, Inc.
May 9, 2025 Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalo... A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent pa... Class II Ortho-Clinical Diagnostics, Inc.
May 9, 2025 Brand Name: VITROS Product Name: VITROS 4600 Chemistry System Model/Catalog... a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent pa... Class II Ortho-Clinical Diagnostics, Inc.
May 9, 2025 Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/C... a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent pa... Class II Ortho-Clinical Diagnostics, Inc.
Apr 14, 2025 VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, o... Product was incorrectly assigned GEN 88 instead of GEN 83 on product carton, foil wrapper, and ca... Class II Ortho-Clinical Diagnostics, Inc.
Mar 24, 2025 Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetell... It was determined that there could be a false positive rate of up to 16.7% for the affected lots ... Class II Ortho-Clinical Diagnostics, Inc.
Dec 19, 2024 Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/... The VITROS 3600 Immunodiagnostic System software versions 3.8.0 may not correctly notify users if... Class II Ortho-Clinical Diagnostics, INc.
Dec 19, 2024 Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalo... The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the S... Class II Ortho-Clinical Diagnostics, INc.
Dec 19, 2024 Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Cat... The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if th... Class II Ortho-Clinical Diagnostics, INc.
Oct 11, 2024 VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively meas... May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, show... Class II Ortho-Clinical Diagnostics, Inc.
Jul 10, 2024 The VITROS XT Chemistry Products ALB-TP Slides is a single device that contai... The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB... Class II Ortho-Clinical Diagnostics, Inc.
Jul 3, 2024 VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2 It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the ... Class II Ortho-Clinical Diagnostics, Inc.
May 21, 2024 VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. ... VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biase... Class II Ortho-Clinical Diagnostics, Inc.
May 14, 2024 VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the V... Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration da... Class III Ortho-Clinical Diagnostics, Inc.
May 8, 2024 VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagno... The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 120... Class III Ortho-Clinical Diagnostics, Inc.
Mar 1, 2024 VITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Ca... Potential for positively biased results due to signal reduction over shelf life. Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS XT 3400 Chemistry System, Catalog No. 6844458 A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 cause... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 680291... A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 cause... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refu... A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC basel... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (ref... A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 cause... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refu... A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 cause... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 680291... A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC basel... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS XT 7600 Integrated System, Catalog No. 6844461 A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 cause... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS XT 3400 Chemistry System, Catalog No. 6844458 A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC basel... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS XT 7600 Integrated System, Catalog No. 6844461 A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC basel... Class II Ortho-Clinical Diagnostics, Inc.
Jan 26, 2024 VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (ref... A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC basel... Class II Ortho-Clinical Diagnostics, Inc.
Jan 22, 2024 VITROS 3600 Immunodiagnostic System (New and Refurbished) Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VI... Class II Ortho-Clinical Diagnostics, Inc.
Jan 22, 2024 VITROS XT7600 Integrated System Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VI... Class II Ortho-Clinical Diagnostics, Inc.
Jan 22, 2024 VITROS 5600 Integrated System (New and Refurbished) Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VI... Class II Ortho-Clinical Diagnostics, Inc.
Nov 27, 2023 VITROS Immunodiagnostic Products Anti-HBs Quantitative Reagent Pack, Catalog ... Affected lots may experience increased calibration failures or an increase in falsely elevated re... Class II Ortho-Clinical Diagnostics, Inc.
Nov 21, 2023 QMS Tacrolimus Calibrators, Product code 10015573 Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry dat... Class II Ortho-Clinical Diagnostics, Inc.
Oct 25, 2023 VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITR... Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coa... Class II Ortho-Clinical Diagnostics, INc.
Oct 10, 2023 VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Prod... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 10, 2023 VITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Co... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 10, 2023 VITROS XT 7600 Integrated System Product Code 6844461 Running Software Versio... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 10, 2023 VITROS 4600 Chemistry System Product Code 6802445 and Refurbished Product Cod... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 10, 2023 VITROS XT 3400 Chemistry System Product Code 6844458 Running Software Version... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Sep 12, 2023 VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 S... Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cau... Class II Ortho-Clinical Diagnostics, INc.
Aug 9, 2023 Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry Sy... Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactur... Class II Ortho-Clinical Diagnostics, Inc.
Aug 5, 2023 VITROS XT Chemistry Products TBIL-ALKP Slides Samples containing high levels of ALKP (or an interfering substance) could deplete the available ... Class II Ortho-Clinical Diagnostics, INc.
Aug 5, 2023 VITROS XT Chemistry Products ALKP Slides Samples containing high levels of ALKP (or an interfering substance) could deplete the available ... Class II Ortho-Clinical Diagnostics, INc.
May 18, 2023 VITROS Immunodiagnostic Products NT-proBNP II Range Verifiers-For in vitro di... May not meet the stability specifications for shelf life outlined in the Instructions For Use (IFU). Class III Ortho-Clinical Diagnostics, Inc.
Apr 17, 2023 VITROS Chemistry Products HbA1c Reagent Kit- IVD for the quantitative determi... Potential for both unexpected positive and negative HbA1c results, may result in misdiagnosis Class II Ortho-Clinical Diagnostics, Inc.

Frequently Asked Questions

When a device safety issue is identified, the manufacturer or FDA initiates a recall. The FDA publishes an enforcement report detailing the device, the reason for the recall, and distribution. Healthcare facilities and providers are notified to check their inventory and take corrective action.

All types — from simple devices like blood glucose test strips and surgical gloves to complex devices like pacemakers, defibrillators, insulin pumps, hip implants, and ventilators. Software-driven devices are increasingly common in recalls due to bugs and cybersecurity vulnerabilities.

Use the search and filter tools on this page. Each recall entry includes the device description, model numbers, and distribution details so you can determine if your device is affected. Contact your healthcare provider if you find a match.