QMS Tacrolimus Calibrators, Product code 10015573
FDA Recall #Z-0712-2024 — Class II — November 21, 2023
Product Description
QMS Tacrolimus Calibrators, Product code 10015573
Reason for Recall
Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it unsuitable for use. Use of the affected lot may results in erroneous calibration results, causing delay in testing.
Recalling Firm
Ortho-Clinical Diagnostics, Inc. — Rochester, NY
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
10 sales units US; 1 sales unit OUS
Distribution
Domestic distribution to IL, IN, FL, NJ, NM, TX. International distribution to Chile.
Code Information
No UDI Lot 74859619, Expiry 11-May-2024
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.