VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
FDA Device Recall #Z-1225-2024 — Class II — January 26, 2024
Recall Summary
| Recall Number | Z-1225-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | January 26, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Ortho-Clinical Diagnostics, Inc. |
| Location | Rochester, NY |
| Product Type | Devices |
| Quantity | 4 US; 18 OUS |
Product Description
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Reason for Recall
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Lot / Code Information
VITROS System Software Version 3.8.1 Regular UDI-DI 10758750012343 Refurbished UD-DI 10758750033201 Serial Numbers US: J46000240 J46000898 J46001030 J46001293 OUS: J46000150 J46001041 J46000190 J46001042 J46000499 J46001186 J46000564 J46001187 J46000801 J46001210 J46000821 J46001217 J46000910 J46001289 J46000933 J46001639 J46001029 J46001645
Other Recalls from Ortho-Clinical Diagnostics, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2559-2025 | Class II | Chemistry Products LAC Slides. Catalog Number: ... | Jul 29, 2025 |
| Z-2560-2025 | Class II | VITROS Chemistry Products LAC Slides. Catalog N... | Jul 29, 2025 |
| Z-2335-2025 | Class II | VITROS Chemistry Products Ca Slides; Catalog N... | Jul 9, 2025 |
| Z-2073-2025 | Class II | VITROS Chemistry Products OP Reagent Gen 15. Mo... | Jun 12, 2025 |
| Z-2121-2025 | Class II | Brand Name: VITROS Chemistry Products Product ... | Jun 10, 2025 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.