VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
FDA Recall #Z-1225-2024 — Class II — January 26, 2024
Product Description
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Reason for Recall
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Recalling Firm
Ortho-Clinical Diagnostics, Inc. — Rochester, NY
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
4 US; 18 OUS
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Code Information
VITROS System Software Version 3.8.1 Regular UDI-DI 10758750012343 Refurbished UD-DI 10758750033201 Serial Numbers US: J46000240 J46000898 J46001030 J46001293 OUS: J46000150 J46001041 J46000190 J46001042 J46000499 J46001186 J46000564 J46001187 J46000801 J46001210 J46000821 J46001217 J46000910 J46001289 J46000933 J46001639 J46001029 J46001645
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.