VITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Calibrator Pack contai...
FDA Recall #Z-2142-2024 — Class II — March 1, 2024
Product Description
VITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Calibrator Pack contains 3 X labelled Calibrator tube (1 per level: C1 - green, C2 - yellow, C3 - red).
Reason for Recall
Potential for positively biased results due to signal reduction over shelf life.
Recalling Firm
Ortho-Clinical Diagnostics, Inc. — Rochester, NY
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
481 units
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MT, NC, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WI, WV and the countries of Bermuda, Colombia, Singapore.
Code Information
Model No. 6801926; GTIN# 10758750001699; Lot 9010 (Manufactured 13-Oct-2023; Expiry 15-Jul-2024) and 9030 (Manufactured 09-Dec-2023; Expiry 26-Aug-2024).
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated