The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin t...

FDA Device Recall #Z-2684-2024 — Class II — July 10, 2024

Recall Summary

Recall Number Z-2684-2024
Classification Class II — Moderate risk
Date Initiated July 10, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Ortho-Clinical Diagnostics, Inc.
Location Rochester, NY
Product Type Devices
Quantity 88253 units

Product Description

The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test side by side separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur, and test results are generated for each analyte independently of the other analyte. The ALB test is a multilayered, analytical element coated on a polyester support. For the albumin measurement, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. When the fluid penetrates the reagent layer, the bromocresol green (BCG) dye diffuses to the spreading layer and binds to albumin in the sample. This binding results in a shift in wavelength of the reflectance maximum of the free dye. The color complex that forms is measured by reflectance spectrophotometry. The amount of albumin-bound dye is proportional to the concentration of albumin in the sample. Model/Catalog Number: 6844293

Reason for Recall

The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT 7600 analyzer. This dust and debris may then settle onto VITROS Na+ (Sodium) Slides as they are processed through the same MicroSlide Subsystem, potentially resulting in an increase in non-reproducible, positively, or negatively biased Na+ results. The level of bias will depend on where the dust and debris settle on the Na+ MicroSlide and may impact Na+ measurements to varying degrees inside and outside of the measuring range depending on the amount of contamination. All of which could vary from no/low bias with no patient harm, to extreme bias (falsely elevated or reduced results) potentially resulting in incorrect diagnosis; treatment; and/or a serious patient health hazard (for example, hypernatremia or hyponatremia). This issue may also potentially generate error codes preventing the analyzer from processing the MicroSlide Assays and cause delay of the patient results. All current and expired VITROS XT Chemistry Products ALB-TP slides are affected by this issue.

Distribution Pattern

Nationwide. Foreign: Australia Belgium Bermuda Brazil Canada L3R 0B8 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden The Netherlands United Kingdom

Lot / Code Information

UDI/GTIN# 10758750031771 All current and expired coatings

Other Recalls from Ortho-Clinical Diagnostics, Inc.

Recall # Classification Product Date
Z-2559-2025 Class II Chemistry Products LAC Slides. Catalog Number: ... Jul 29, 2025
Z-2560-2025 Class II VITROS Chemistry Products LAC Slides. Catalog N... Jul 29, 2025
Z-2335-2025 Class II VITROS Chemistry Products Ca Slides; Catalog N... Jul 9, 2025
Z-2073-2025 Class II VITROS Chemistry Products OP Reagent Gen 15. Mo... Jun 12, 2025
Z-2121-2025 Class II Brand Name: VITROS Chemistry Products Product ... Jun 10, 2025

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.