DateProductReasonClassFirm
Oct 16, 2012 bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML01, Clinical instrument s... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 16, 2012 bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML 02, Clinical instrument ... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 16, 2012 bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML05, Clinical instrument s... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 16, 2012 bioMerieux Kit, Instrument & Update, B.40, 3D I ZIP-ML02, Industry instrument... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 16, 2012 bioMerieux Kit, Instrument & Update, B.40, 3D I USB-ML01, Industry instrument... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 16, 2012 bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML04, Clinical instrument s... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 16, 2012 bioMerieux Kit, Instrument & Update, 3D B.40 USB-ML05, Clinical instrument so... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 16, 2012 bioMerieux Kit, Instrument & Update, 3D IND USB-ML 02, Industry instrument so... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 16, 2012 bioMerieux Kit, Instrument & Update, B.40 Dualt-ML, Industry instrument softw... The product may mislabel patient information on culture specimen bottles due to a software miscom... Class II Biomerieux Inc
Oct 11, 2012 COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Auto... Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection... Class II Roche Molecular Systems, Inc.
Oct 9, 2012 Digital RID Plate Reader and Software Product Code: AD400 If a control ring is marked after reading, the software will not flag results that are out of the... Class II The Binding Site Group, Ltd.
Oct 5, 2012 Pinnacle3 VCC P/N 4598 001 41341 REV A, Philips Medical Systems (Cleveland), ... Philips Medical Systems Cleveland), Inc. is recalling Pinnacle3 Virtual Client Connection softwa... Class II Philips Medical Systems (Cleveland), Inc.
Oct 2, 2012 Centricity Cardio Imaging 5.0 is a software only Picture Archiving and Commun... GE Healthcare is aware of a potential safety risk where an image labeling situation exists in the... Class II GE Healthcare It
Sep 28, 2012 Computed Tomography X-Ray System, Extended Brilliance Workspace (EBW), Model ... Philips is issued a Field Change Order to update Extended Brilliance Workspace (EBW) systems to s... Class II Philips Medical Systems (Cleveland) Inc
Sep 28, 2012 Philips Medical System Extended Brilliance Workspace (EBW), Computed Tomograp... Incorrect values may be displayed. Philips is issuing a software update (version 4.5.5) to the E... Class II Philips Medical Systems (Cleveland) Inc
Sep 21, 2012 Diamigo i-Phone App The Diamigo app was intended for use as an educationa... Diamigo, a software application, was released through the i-Phone Global Store which inadvertentl... Class II sanofi-aventis US, Inc.
Sep 20, 2012 ABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2 Horiba Medical is recalling ABX PENTRA ML because they have confirmed an issue with the ABX PENTR... Class II Horiba Instruments, Inc dba Horiba Medical
Sep 14, 2012 Hologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4... InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical ... Class II Hologic, Inc.
Sep 14, 2012 Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1an... Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may b... Class II Hologic, Inc.
Sep 12, 2012 The RIO (TGS 2.), Model # MAKO TGS 2.0 (Part No. 204000). The RIO is inten... MAKO Surgical Group recalled their RIO System software, version 2.4 and is implementing software ... Class II Mako Surgical Corporation
Sep 11, 2012 Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 P... Philips was notified that when performing post processing with the Brilliance Workspace Portal pl... Class II Philips Medical Systems (Cleveland) Inc
Sep 3, 2012 UniCel¿ DxC 600/800 with Software v5.0.11, Part Number B12218. The UniCel... Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment proce... Class II Beckman Coulter Inc.
Aug 28, 2012 The Brilliance CT 16 System Product Usage: The Brilliance 16 is a Whole B... Philips was notified that the system logout in software version 2.3.6 is now longer and may somet... Class III Philips Medical Systems (Cleveland) Inc
Aug 24, 2012 Volcano brand Intravascular Ultrasound, Volcano s5/s5i Family Software Versio... The display on the system has the potential to change. In house testing eventually identified the... Class II Volcano Corporation
Aug 21, 2012 Oncentra External Beam 4.1. Radiation treatment planning software designed... During the planning process the beam weights can be changed by the customers. When the beam weigh... Class II Elekta, Inc.
Aug 17, 2012 Octopus 101 Perimeter; PeriTrend Software Analysis A perimeter is a device... The firm recalled the device after learning of a possible deviation in the stimulus luminance of ... Class II Haag-Streit USA Inc
Aug 16, 2012 Zoll X Series Defibrillator/Pacemaker/Monitor Product Usage: The product ... Shipped with incorrect software Class II ZOLL Medical Corporation
Aug 16, 2012 GE Healthcare, Innova 2121IQ, Innova 3131IQ Cardiovascular X-ray imaging syst... GE Healthcare has recently become aware of a potential safety issue with respect to the Innova 21... Class II GE Healthcare, LLC
Aug 13, 2012 GE Healthcare, LOGIQ S8 Ultrasound System The device is intended for use ... GE Healthcare became aware of a potential issue related to incorrect image measurement associated... Class II GE Healthcare, LLC
Aug 13, 2012 Software for Vitrea, Vitrea fX, and Vitrea Enterprise Suite (VES): VPMC... Vital Images has found two potential errors in calcium score values in restored snapshots of stud... Class II Vital Images, Inc.
Aug 8, 2012 Positron Emission Tomography (PET) and X-ray Computed Tomography, GEMINI TF 1... Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a... Class II Philips Medical Systems (Cleveland) Inc
Aug 3, 2012 GE Healthcare Brivo XR385 DR-F/ X-Ray System X-Ray System GE Healthcare has become aware of a software issue on the interface of the Brivo XR385 DR-F/ Mode... Class II GE Healthcare, LLC
Aug 1, 2012 GE Healthcare Optima XR 220amx 15kw Mobile X-Ray System The GE Automatic M... GE Healthcare has become aware of a software issue on the interface of the mobile radiographic pr... Class II GE Healthcare, LLC
Aug 1, 2012 GE Healthcare Optima XR 200 amx with Digital Upgrade. The GE Automatic Mob... GE Healthcare has become aware of a software issue on the interface of the mobile radiographic pr... Class II GE Healthcare, LLC
Aug 1, 2012 GE Healthcare Optima XR 220amx 30kw Mobile X-ray System The GE Automatic M... GE Healthcare has become aware of a software issue on the interface of the mobile radiographic pr... Class II GE Healthcare, LLC
Jul 30, 2012 DEXIS Imaging Suite - catalog number 1.008.7932, version numbers 10.0.0 throu... A condition can occur during renumbering of the patient database using the DEXCopy feature. When... Class II Imaging Sciences International, LLC
Jul 24, 2012 Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended ... Customers who received a Capintec Captus 3000 Thyroid Uptake System with Software Revisions 1.22,... Class II Capintec Inc
Jul 23, 2012 VersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek Sys... Review of instrument software databases and investigation of two customer reports determined that... Class III Trek Diagnostic Systems
Jul 23, 2012 Cisco brand HealthPresence, version 2.1; Model/Catalog Number: CHP-ENTPRSYR2.... Software anomaly may result in incorrect values and interpretations. The initial vital signs read... Class II Cisco Systems Inc
Jul 18, 2012 The Brilliance 64 and Ingenuity CT are whole body Computed Tomography X-Ray S... 1. The current system design of software version 4.0.0xx379 does not prevent the system from scan... Class II Philips Medical Systems (Cleveland) Inc
Jul 16, 2012 DataCara GM is an information management system that provides data management... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 The COBAS 8000 data manager modular analyzer series (all versions) is a is a... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 The COBAS INTEGRA 400/400 plus system (software version 3.4) is a fully autom... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 Data Station AMPLILINK software is an instrument user interface/ data managem... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 DataInnovations (Instrument Manager) (version 8.04) interfaces middleware bet... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 The MagNA Pure LC (version 3.0) is a sample preparation instrument that can... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 LightCycler 2.0 Instrument (software version 4.1) is a Rapid PCR analyzer wit... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 LightCycler 1.2 (software version 3.5) is a Rapid PCR analyzer with real-time... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 The MagNA Pure 96 System (versions 1.0 and 2.0) is a sample preparation ins... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.
Jul 16, 2012 DataCare POC  is an information management system that provides data managem... Roche Molecular Systems, Inc. has become aware of a device-associated diagnostic software issue. Class II Roche Molecular Systems, Inc.