Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by traine...
FDA Device Recall #Z-2312-2012 — Class II — July 24, 2012
Recall Summary
| Recall Number | Z-2312-2012 |
| Classification | Class II — Moderate risk |
| Date Initiated | July 24, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Capintec Inc |
| Location | Pittsburgh, PA |
| Product Type | Devices |
| Quantity | 256 |
Product Description
Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform thyroid uptake procedures, which consist of the measurement of organ uptake and circulating radioactivity.
Reason for Recall
Customers who received a Capintec Captus 3000 Thyroid Uptake System with Software Revisions 1.22, 1.23, 1.24, 1.25, and 1.26 may encounter an error while using the thyroid uptake protocol ?Measure Liquid and Subtract Residual." This protocol was a new feature added to Software Revision 1.22 in January 2011. The error occurs only if the user does NOT measure the residual dose immediately after me
Distribution Pattern
Worldwide distribution: USA (nationwide) and countries of: Argentina, Belgium, Brazil, Canada, China, Greece, Hong Kong, India, Israel, Malaysia, Philippines, Spain, South Korea, Thailand, and Turkey.
Lot / Code Information
Catalog numbers for new systems: 5430-0076, 5430-0077, 5430-0103, and 5430-0087. Catalog numbers for conversion systems: 5430-0118 and 5430-0119. Serial numbers for new systems: 901006 through 901394, inclusive. Serial numbers for conversion systems: CNV 674 through CNV 730, inclusive. Also affected are devices which were serviced and devices for which customers received software upgrades between January 20, 2011 and June 30, 2012 with Revision 1.22, 1.23, 1.24, 1.25, or 1.26. (NOTE: Item number 5430-0075, Captus 3000 well counter system, is excluded. This system does not have the ability to perform thyroid uptake procedures.)
Other Recalls from Capintec Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1564-2020 | Class II | Captus 4000e Thyroid Uptake System, Model Numbe... | Mar 3, 2020 |
| Z-0551-2019 | Class II | Captus 700 Thyroid Uptake System, Model # 5430-... | Nov 5, 2018 |
| Z-0550-2019 | Class II | Captus 3000 Thyroid Uptake System, Model # 5430... | Nov 5, 2018 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.