Software for Vitrea, Vitrea fX, and Vitrea Enterprise Suite (VES): VPMC-09293C Vitrea . Mod...

FDA Device Recall #Z-0186-2013 — Class II — August 13, 2012

Recall Summary

Recall Number Z-0186-2013
Classification Class II — Moderate risk
Date Initiated August 13, 2012
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Vital Images, Inc.
Location Plymouth, MN
Product Type Devices
Quantity 2088 (1133 USA)

Product Description

Software for Vitrea, Vitrea fX, and Vitrea Enterprise Suite (VES): VPMC-09293C Vitrea . Model Number Vitrea 5.2, 6.0, 6.1, and 6.2 VPMC-09346B Vitrea fX Model Number Vitrea fX 3.1, 5.2, 6.0, 6.1, and 6.2 VPMC-11731A Vitrea Enterprise Suite Model Number Vitrea Enterprise Suite 1.3, 6.0, 6.1, and 6.2. The software is a medical diagnostic system that allows the processing, review, analysis, communication, and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. It provides the ability to review digital images from original DICOM data for multiple modalities, including CT, MR, DR, CR, XA, US, NM, PET, etc. The software is not meant for primary image interpretation in mammography. The Vitrea VScore option is intended for cardiac scoring from whole body CT derived measurements, including non-invasive detection and quantification of atherosclerotic plaque. Two image processing options, EKG Gate and Auto Gate, allow the operator to select images with reduced motion artifacts when processing data for Coronary Artery Calcification Scoring.

Reason for Recall

Vital Images has found two potential errors in calcium score values in restored snapshots of studies generated through use of Vitrea or Vitrea Enterprise Suite. This issue affects Version 3.1 of Vitrea fX; Version 5.2 of Vitrea; Versions 6.0, 6.1 and 6.2 of Vitrea and Vitrea fX; and Versions 1.3, 6.0, 6.1 and 6.2 of Vitrea Enterprise Suite; and all updates to those versions.

Distribution Pattern

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, and the countries of ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BELARUS, BELGIUM, BRAZIL, BULGARIA, CANADA CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, FINLAND, FRANCE, GREAT BRITAIN, GEORGIA, GERMANY, GHANA, GREECE, GUAM, HONDURAS, HONG KONG, INDIA, INDONESIA, IRAQ, IRELAND, ITALY, IVORY COAST, JAPAN, JORDAN KAZAKHSTAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MEXICO, MALI, MOSCOW, MYANMAR, NAMIBIA, NETHERLANDS, NEW ZEALAND, NIGERIA, NORTHERN IRELAND, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SENEGAL, SERBIA, SERBIA AND MONTENEGRO, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SUDAN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, THE NETHERLANDS, TIBET, TURKEY, UKRAINE, UNITED KINGDOM, UZBEKISTAN, VENEZUELA, VIETNAM, AND YEMEN

Lot / Code Information

Model Number Vitrea 5.2, 6.0, 6.1, and 6.2 VPMC-09346B Vitrea fX Model Number Vitrea fX 3.1, 5.2, 6.0, 6.1, and 6.2 VPMC-11731A Vitrea Enterprise Suite Model Number Vitrea Enterprise Suite 1.3, 6.0, 6.1, and 6.2.

Other Recalls from Vital Images, Inc.

Recall # Classification Product Date
Z-0169-2014 Class II VitreaCore¿ software. Versions subject to the l... Oct 21, 2013
Z-1181-2013 Class II Vitrea¿ CT Brain Perfusion made available in va... Mar 15, 2013
Z-1144-2013 Class II Vitrea CT Multi-Chamber Cardiac Functional Anal... Mar 15, 2013
Z-1002-2013 Class II Vitrea Enterprise Suite, Vitrea, VitreaAdvanced... Feb 5, 2013

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.