| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| May 18, 2016 | VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Uniqu... | Ortho Clinical Diagnostics (OCD) confirmed a potential for biased calcium (Ca) results to be gene... | Class II | Ortho-Clinical Diagnostics |
| May 16, 2016 | enGen Laboratory Automation System, Product Code engen, Software Versions enG... | If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numeri... | Class III | Ortho-Clinical Diagnostics |
| May 9, 2016 | VITROS Immunodiagnostic Products AFP Reagent Pack, REF/Product Code 680 0784,... | Ortho Clinical Diagnostics (OCD) confirmed that specific lots of VITROS Immunodiagnostic Products... | Class III | ORTHO-CLINICAL DIAGNOSTICS |
| Apr 19, 2016 | The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunctio... | There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will pr... | Class III | Ortho-Clinical Diagnostics |
| Apr 6, 2016 | VITROS 3600 Immunodiagnostic System, Catalog Number 6802783,for use in the in... | Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 ... | Class II | Ortho-Clinical Diagnostics |
| Apr 6, 2016 | VITROS 5600 Integrated System, Catalog Number 6802413, Unique Device Identifi... | Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 ... | Class II | Ortho-Clinical Diagnostics |
| Mar 17, 2016 | enGen (TM) Laboratory Automation System configured with Thermo Scientific Rec... | The electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may l... | Class II | Ortho-Clinical Diagnostics |
| Feb 18, 2016 | 1) VITROS¿ Immunodiagnostic Products Total Beta-hCG II Reagent Pack (Unique ... | When testing was conducted using certain lots, VITROS¿ Systems generated results within the measu... | Class II | Ortho-Clinical Diagnostics |
| Feb 16, 2016 | VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & ... | The two sets of wires (4 total wires) exiting from the EMI filters were switched, creating a situ... | Class II | Ortho-Clinical Diagnostics |
| Dec 21, 2015 | VITROS 4600 Chemistry System, Catalog Number 6802445, Unique Device Identifie... | Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chem... | Class II | Ortho-Clinical Diagnostics |
| Dec 21, 2015 | VITROS 250 Chemistry Systems, Catalog 8132086, Unique Device Identifier No. 1... | Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chem... | Class II | Ortho-Clinical Diagnostics |
| Dec 21, 2015 | VITROS 350 Chemistry System, Catalog Number 6802153, Unique Device Identifier... | Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chem... | Class II | Ortho-Clinical Diagnostics |
| Dec 21, 2015 | VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identif... | Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chem... | Class II | Ortho-Clinical Diagnostics |
| Dec 21, 2015 | VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifie... | Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chem... | Class II | Ortho-Clinical Diagnostics |
| Nov 3, 2015 | VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifie... | Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reage... | Class II | Ortho-Clinical Diagnostics |
| Nov 3, 2015 | VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifie... | Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reage... | Class II | Ortho-Clinical Diagnostics |
| Nov 3, 2015 | VITROS 4600 Chemistry System, Catalog Number 6802445, Unique Device Identifie... | Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reage... | Class II | Ortho-Clinical Diagnostics |
| Nov 3, 2015 | VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identif... | Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reage... | Class II | Ortho-Clinical Diagnostics |
| Sep 28, 2015 | VITROS Chemistry Products Calibrator Kit 9 Calibrator, 2 mL, Product Code 85... | The firm is conducting a product correction due to an increase in U90-382 or 6LU condition codes ... | Class III | Ortho-Clinical Diagnostics |
| Sep 10, 2015 | VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique D... | Software Anomaly during ADD Installation on VITROS 5,1 FS Chemistry Systems using Software Versio... | Class II | Ortho-Clinical Diagnostics |
| Jun 30, 2015 | VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique D... | Software anomaly may occur during the installation of an Assay Data Diskette (ADD) using Software... | Class II | Ortho-Clinical Diagnostics |
| May 6, 2015 | enGen Laboratory Automation System (enGen System), Product Code ENGEN, config... | Recapper Caps may fall off of, or be knocked from sample tubes placed in the storage racks of ent... | Class II | Ortho-Clinical Diagnostics |
| May 6, 2015 | VITROS Immunodiagnostic Products Signal Reagent, Product Code 1072693, Unique... | The Immunodiagnostic Products Signal Reagent, Lot 3082 as the reagent packs may contain blank lab... | Class III | ORTHO-CLINICAL DIAGNOSTICS |
| Apr 6, 2015 | VITROS 5600 Chemistry System, Software Version 3.2 & Below. Automates pre-... | Software anomaly allows testing of multiple assays using a single urine specimen, regardless of t... | Class II | Ortho-Clinical Diagnostics |
| Apr 6, 2015 | VITROS 4600 Chemistry System, Software Version 3.2 & Below. Automates pre-... | Software anomaly allows testing of multiple assays using a single urine specimen, regardless of t... | Class II | Ortho-Clinical Diagnostics |
| Apr 6, 2015 | Assay Data Disk (ADD), All DRV versions that support product within expiry da... | Software anomaly regarding urine samples that require acidified pretreatment. The software has a... | Class II | Ortho-Clinical Diagnostics |
| Apr 6, 2015 | VITROS 5,1 FS Chemistry System, Software Version 2.8 & Below Automates pre... | Software anomaly regarding urine samples that require acidified pretreatment. The software has a... | Class II | Ortho-Clinical Diagnostics |
| Mar 10, 2015 | Software Version 2.8 & Below on VITROS 5,1 FS Chemistry Systems, Catalog Numb... | Software Anomaly; It is possible for the device to process samples with a cartridge other than th... | Class II | Ortho-Clinical Diagnostics |
| Dec 29, 2014 | VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the fo... | In combination with the Batch Programming option (only) for programming samples, the hand-held ba... | Class II | Ortho-Clinical Diagnostics |
| Nov 19, 2014 | VITROS 5600 Integrated System, Catalog Number 6802413, IVD --- Ortho Clinical... | Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and ... | Class II | Ortho-Clinical Diagnostics |
| Nov 19, 2014 | VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Catalog Nu... | Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and ... | Class II | Ortho-Clinical Diagnostics |
| Nov 19, 2014 | VITROS 3600 Immunodiagnostic System, Catalog Number 6802783, IVD --- Ortho C... | Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and ... | Class II | Ortho-Clinical Diagnostics |
| Nov 6, 2014 | VITROS CHEMISTRY PRODUCTS TRIG Slides, REF 133 6544, IVD For in vitro di... | Potential for biased results using the specific lot of VITROS CHEMISTRY PRODUCTS TRIG slides. | Class II | Ortho-Clinical Diagnostics |
| Sep 29, 2014 | VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemis... | Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Prod... | Class III | Ortho-Clinical Diagnostics |
| Sep 8, 2014 | VITROS 5600 Integrated System, System Product Code 6802413. For use in the... | Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the... | Class II | Ortho-Clinical Diagnostics |
| Sep 8, 2014 | VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), System Pro... | Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the... | Class II | Ortho-Clinical Diagnostics |
| Aug 20, 2014 | VITROS Chemistry Products PHYT (Phenytoin) Slides, REF/Product Code 829 8671,... | Ortho Clinical Diagnostics identified a potential for biased results to be generated when using V... | Class II | Ortho-Clinical Diagnostics |
| Aug 6, 2014 | VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitati... | VITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower... | Class II | ORTHO-CLINICAL DIAGNOSTICS |
| Jul 16, 2014 | VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot ... | Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 1... | Class II | ORTHO-CLINICAL DIAGNOSTICS |
| Jul 15, 2014 | VITROS¿ Chemistry Products CDM PROM Calibration Data Module Rev. No. 0168 and... | Inability to calibrate or use VITROS NH3 DT Slides, GEN 59 with VITROS Chemistry Products DT Cali... | Class II | Ortho-Clinical Diagnostics |
| Jul 7, 2014 | VITROS 5600 Integrated System --- For use in the in vitro quantitative, semi-... | Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and ... | Class II | Ortho-Clinical Diagnostics |
| Jul 7, 2014 | VITROS 4600 Chemistry System (VITROS 5,1 FS System family member) --- The VIT... | Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and ... | Class II | Ortho-Clinical Diagnostics |
| Jul 7, 2014 | VITROS 3600 Immunodiagnostic System --- For use in the in vitro quantitative,... | Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and ... | Class II | Ortho-Clinical Diagnostics |
| May 12, 2014 | VITROS (R) Chemistry Products FS Calibrator 1, Product Code/ Catalog Number 6... | Calibrator values assigned for the calibrator kit lots were released using a reference method tha... | Class II | Ortho-Clinical Diagnostics |
| May 12, 2014 | VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number... | Calibrator values assigned for the calibrator kit lots were released using a reference method tha... | Class II | Ortho-Clinical Diagnostics |
| Apr 17, 2014 | enGen (TM) Laboratory Automation System Configured with TCAutomation" Softwar... | Software anomaly: potential buffer timeout issues when using enGen (TM) Laboratory Automation Sys... | Class II | Ortho-Clinical Diagnostics |
| Apr 10, 2014 | VITROS 4600 Chemistry System, Product Code/Catalog Number 6802445 for in v... | Potential for biased results when a Multiple Window (MW) code is associated with results generate... | Class II | Ortho-Clinical Diagnostics |
| Apr 10, 2014 | VITROS 5600 Integrated System, Product Code/Catalog Number 6802413 for in vit... | Potential for biased results when a Multiple Window (MW) code is associated with results generate... | Class II | Ortho-Clinical Diagnostics |
| Apr 10, 2014 | enGen Laboratory Automation Systems (enGen) with TCAutomation -- For in vit... | Software anomaly: potential for delay in the reporting of patient sample test results when using ... | Class II | Ortho-Clinical Diagnostics |
| Apr 10, 2014 | VITROS 5,1 FS Chemistry System and VITROS 5,1 FS Refurbished, Product Codes/... | Potential for biased results when a Multiple Window (MW) code is associated with results generate... | Class II | Ortho-Clinical Diagnostics |