enGen Laboratory Automation System (enGen System), Product Code ENGEN, configured with all of the...
FDA Device Recall #Z-1724-2015 — Class II — May 6, 2015
Recall Summary
| Recall Number | Z-1724-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 6, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Ortho-Clinical Diagnostics |
| Location | Rochester, NY |
| Product Type | Devices |
| Quantity | 45 units worldwide (Domestic: 19 units, Foreign: 26) |
Product Description
enGen Laboratory Automation System (enGen System), Product Code ENGEN, configured with all of the following system components: Recapper Caps (Product Codes 952238 or 6844286 only), Recapper Module (230V) or Recapper Module (110V) (Product Codes 952036-EG or 952136-EG), Rack Entry and Exit Module and/or Rack Exit Module (Product Codes 952020-EG/6844086 or 952019-EG/6844103), Storage Rack Slim 100 positions or Storage Rack 180 positions (Product Codes 952075 or 952073). COMMON NAME: enGen Track System. The enGen Lab Automation System is a modification to VITROS Analyzers that includes the addition of an automated track sub-system (TCAutomation) for pre-analytical and post-analytical sample and data management.
Reason for Recall
Recapper Caps may fall off of, or be knocked from sample tubes placed in the storage racks of entry/exit modules. The gripper may grab a tube at the wrong height and drop the tube in the Rack Entry/Exit or Rack Exit module. These situations may result in splashing of bio-hazardous fluid. Sample cross-contamination may occur with an uncapped tube in the rack.
Distribution Pattern
Worldwide distribution. US Nationwide, Australia, Bermuda, Brazil, Canada, Chile, China, France, India, Italy, Norway, Portugal, Spain, Sweden, and the United Kingdom.
Lot / Code Information
Serial Numbers: J89134, JIM341226, 341516, JIM342964, JIM340432, JIM340814, IM341243, JIM336912, J271265, J271369, J271372, J271267, J85208, JIM342217, J7023890, J7023841, J7025007, J7024894, J7023648, J7021530, J57662, J8006290, J75685, J30370, J3055510, J3055507, JIM341820, JIM341679, JIM341174, JIM340667, JIM-341592, JIM336957, JIM341477, JIM336855, JIM334236, JIM341415, JIM341392, JIM340934, JIM336167, JIM342249, JIM341750, JIM341426, JIM334527, JIM341021, JIM340658.
Other Recalls from Ortho-Clinical Diagnostics
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1372-2024 | Class II | VITROS Performance Verifier I | Feb 27, 2024 |
| Z-1373-2024 | Class II | VITROS Performance Verifier II | Feb 27, 2024 |
| Z-2358-2023 | Class II | VITROS Immunodiagnostic Products Troponin I ES ... | Jun 5, 2023 |
| Z-2357-2023 | Class II | VITROS Immunodiagnostic Products Troponin I ES ... | Jun 5, 2023 |
| Z-2129-2023 | Class II | VITROS Chemistry Products Calibrator Kit 20-in ... | May 8, 2023 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.