enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V o...

FDA Device Recall #Z-1569-2016 — Class II — March 17, 2016

Recall Summary

Recall Number Z-1569-2016
Classification Class II — Moderate risk
Date Initiated March 17, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Ortho-Clinical Diagnostics
Location Rochester, NY
Product Type Devices
Quantity US: 19, Foreign: 30

Product Description

enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL NAME: enGen (TM) Track System -- NOTE: Outside the US the system is also known as VITROS Automation Solution (VAS) TC Automation System (configured with Thermo Scientific Recapper Module from VITROS Automation Solutions), Product Code 6844300 --- The enGen (TM) Lab Automation System is a modification to VITROS Analyzers that includes the addition of an automated track sub-system (TC Automation) for pre-analytical and post-analytical sample and data management. In addition to an Automated Track Sub-system, the enGen (TM) System includes one or more VITROS Chemistry, Immunodiagnostic and/or Integrated Systems (clinical analyzers) as well as a Data Management Subsystem (Middleware), and a Software Configuration (gsb) File. The enGen System is typically interfaced with a Laboratory Information System (LIS) via connectivity facilitated at the Data Management Subsystem level. The enGen (TM) Laboratory Automation System contains a module (i.e. enGen (TM)Laboratory Automation Recapper Module) that applies caps to the sample collection container after the sample has been aspirated. In this module, an electrical cable supplies power to the cap vibratory feeder and is supposed to be located above the vibratory feeder support plate. The cap vibratory feeder sits upon a feeder support plate. Un-capped sample containers pass underneath this support plate prior to being re-capped for storage. The Recapper Module is supplied by Thermo Fisher Scientific.

Reason for Recall

The electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may loosen from its correct position above the vibration feeder support plate and migrate to under the support plate. If this occurs, the uncapped sample tubes may hit the cable, and sample fluid (e.g., serum) may splash from the tube and into another tube or onto the under carriage of the support plate.

Distribution Pattern

Distributed in the states of MD, NY, ID, NC, CO, OH, PA, TN, MS, MO, CA, DE, FL, GA, AZ, NJ, and WV and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, France, India, Italy, Norway, Portugal, Saudi Arabia, Spain, Sweden, Thailand, and United Kingdom.

Lot / Code Information

US Serial Numbers: JFBX395J, GBX395J, J953252, JJXBQW4J, J3RHC23J, J86PWC5J, J5LQ6Q2J, JFCM073J, J54FC01J, J5J4B45J, J7KSC15J, J4HX7S4J, JGWXQ52J, J953221, JG7BFB5J, JHPKG25J, JCXL941J, J736BM4J, JFXCBG3J; -- Foreign Serial Numbers: J89133 (VAS), JHWJZY4J, J3J4B45J, J953244, JCT20Q3J, J16XQH4J, GWJVY4J, DWX052J, J271264, J271368, J271371, J271269, J85206, J953216, DP2LF62, J7023874, J7023847, J7025017, J7024893, J7023645, J7023435, J57663, J800103, S/N 89VQ762 (VAS), J75863, J75691, J30372, J3055503, J3055496.

Other Recalls from Ortho-Clinical Diagnostics

Recall # Classification Product Date
Z-1372-2024 Class II VITROS Performance Verifier I Feb 27, 2024
Z-1373-2024 Class II VITROS Performance Verifier II Feb 27, 2024
Z-2358-2023 Class II VITROS Immunodiagnostic Products Troponin I ES ... Jun 5, 2023
Z-2357-2023 Class II VITROS Immunodiagnostic Products Troponin I ES ... Jun 5, 2023
Z-2129-2023 Class II VITROS Chemistry Products Calibrator Kit 20-in ... May 8, 2023

Frequently Asked Questions

Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.