VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750002...
FDA Recall #Z-1074-2016 — Class II — December 21, 2015
Product Description
VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750002740; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Reason for Recall
Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).
Recalling Firm
Ortho-Clinical Diagnostics — Rochester, NY
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
US = 1009; Foreign = 1003
Distribution
Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela
Code Information
Software Version 3.2 and below; Serial Numbers J56000024 - 56002339
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.