VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Syste...

FDA Recall #Z-1263-2015 — Class II — December 29, 2014

Recall #Z-1263-2015 Date: December 29, 2014 Classification: Class II Status: Terminated

Product Description

VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemistry System, VITROS 3600 Immunodiagnostics System, VITROS 4600 Chemistry System, VITROS 5600 Integrated System. This product is an accessory for use with VITROS Analyzer Systems.

Reason for Recall

In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended.

Recalling Firm

Ortho-Clinical Diagnostics — Rochester, NY

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

Domestic: 62 units; Foreign: 10 units

Distribution

Worldwide Distribution-US (nationwide) and the countries of Canada, Australia and England.

Code Information

Lot Numbers: 0914F11968, 0914F11969, 0914F11970, 0914F11971, 0914F11972, 0914F11973, 0914F11974, 0914F11975, 0914F11976, 0914F11977, 0914F11978, 0914F11979, 0914F11980, 0914F11981, 0914F11982, 0914F11983, 0914F11984, 0914F11985, 0914F11986, 0914F11987, 0914F11988, 0914F11989, 0914F11990, 0914F11997, 0914F12007, 0914F12016, 0914F12017, 0914F12020, 0914F12022, 0914F12028, 0914F12033, 0914F12037, 0914F12039, 0914F12040, 0914F12041, 0914F12046, 0914F12048, 0914F12049, 0914F12050, 0914F12053, 0914F12054, 0914F12055, 0914F12056, 0914F12057, 0914F12058, 0914F12059, 0914F12060, 0914F12061, 0914F12062, 0914F12063, 0914F12064, 0914F12065, 0914F12066, 0914F12067, 0914F12081, 0914F12082, 0914F12083, 0914F12084, 0914F12085, 0914F12086, 1014F12318, 1014F12319, 1014F12320, 1014F12321, 1014F12322, 1014F12323, 1014F12386, 1014F12387, 1014F12388, 1014F12389, 1014F12390, 1014F12391.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated