| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Mar 30, 2022 | Dragonfly OpStar Imaging Catheter, REF 1014651/1014652, 0.068 in, 0.014 in, ... | There is a potential that the proximal marker may separate from the imaging catheter | Class I | Abbott Vascular |
| Mar 30, 2022 | The VerifyNow P2Y12 Test is designed to measure platelet P2Y12 receptor block... | Two whole blood Platelet Reactivity tests exist that share the same reagent formulation and are u... | Class II | Accriva Diagnostics, Inc. |
| Mar 25, 2022 | BD Anti-Kappa APC (Cat. No. 341098) is an analyte specific reagent for the An... | Anti-Kappa APC & Anti-Kappa PE analyte reagents may exhibit uncharacteristic flow profiles. | Class II | Becton, Dickinson and Company, BD Biosciences |
| Mar 25, 2022 | BD Anti-Kappa PE (Cat. No. 346587) is an analyte specific reagent for the Ant... | Anti-Kappa APC & Anti-Kappa PE analyte reagents may exhibit uncharacteristic flow profiles. | Class II | Becton, Dickinson and Company, BD Biosciences |
| Mar 23, 2022 | TearCare version 1.0 SmartHubs | Ophthalmic devices distributed prior to the recently-obtained 510(k) clearance do not fall within... | Class II | Sight Sciences, Inc. |
| Mar 23, 2022 | Medtronic IN.PACT Admiral, Paclitaxel-coated PTA Balloon Catheter, Sterile EO... | Packaging for Balloon Catheters may be damaged resulting in loss of sterility. | Class II | Medtronic Vascular, Inc. |
| Mar 23, 2022 | Medtronic IN.PACT AV, Paclitaxel-coated PTA Balloon Catheter, REF: IAV0600800... | Packaging for Balloon Catheters may be damaged resulting in loss of sterility. | Class II | Medtronic Vascular, Inc. |
| Mar 18, 2022 | Panther Fusion GBS Assay Cartridges 96 Tests, Catalog Number: PRD-04484. in ... | Samples, run on a GBS Assay, that have analyte concentrations that are at or near the limit of de... | Class II | Hologic, Inc |
| Mar 17, 2022 | da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Su... | As a result of complaint data review, it identified tissue pushout events occurring during firing... | Class II | Intuitive Surgical, Inc. |
| Mar 11, 2022 | 20/30 Priority Pack w/COPILOT - REF 1003327 WPL2122272-01 (2019-03-01) | Due to an increase in complaint trend for leaks and intermittent/loose connections. | Class II | Abbott Vascular |
| Mar 11, 2022 | 20/30 Priority Pack Accessory Kit/.096 RHV - REF 1000186 WPL2122270-01 (2019... | Due to an increase in complaint trend for leaks and intermittent/loose connections. | Class II | Abbott Vascular |
| Mar 11, 2022 | Total Resources International First Aid Kits, Item Numbers: 10HBC01087, 10HBC... | ULTRAtab Laboratories recalled Acetaminophen and electrolyte tablets that were later placed into ... | Class II | Total Resources Intl |
| Mar 11, 2022 | 20/30 INDEFLATOR -REF 1000184 WPL2122268-01 (2019-03-01) | Due to an increase in complaint trend for leaks and intermittent/loose connections. | Class II | Abbott Vascular |
| Mar 11, 2022 | Total Resources International First Aid Kits, Item Numbers: 01REF2621, 10OUT0... | ULTRAtab Laboratories recalled Acetaminophen and electrolyte tablets that were later placed into ... | Class II | Total Resources Intl |
| Mar 11, 2022 | INDEFLATOR Plus 30 - REF 1000183 WPL2122277-01 (2019-03-01) | Due to an increase in complaint trend for leaks and intermittent/loose connections. | Class II | Abbott Vascular |
| Mar 11, 2022 | Plus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Pl... | Due to an increase in complaint trend for leaks and intermittent/loose connections. | Class II | Abbott Vascular |
| Mar 10, 2022 | Gallant VR Implantable Cardioverter Defibrillator REF # CDVRA500Q | Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to e... | Class II | St. Jude Medical, Cardiac Rhythm Management Div... |
| Mar 10, 2022 | Merlin PCS 3650 programmer Model 3330 software | Due to a programmer software anomaly under very specific circumstance when executing a pacing cap... | Class II | St. Jude Medical, Cardiac Rhythm Management Div... |
| Mar 10, 2022 | Merlin 2 PCS MER3700 programmer Model MER3400 software | Due to a programmer software anomaly under very specific circumstance when executing a pacing cap... | Class II | St. Jude Medical, Cardiac Rhythm Management Div... |
| Mar 10, 2022 | Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q | Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to e... | Class II | St. Jude Medical, Cardiac Rhythm Management Div... |
| Mar 10, 2022 | Gallant HF Implantable Cardioverter Defibrillator REF# CDHFA500Q | Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to e... | Class II | St. Jude Medical, Cardiac Rhythm Management Div... |
| Mar 4, 2022 | First SIGN SARS-CoV-2 Antigen Test | Frm became aware that a customer complaint was reported to US FDA that SARS-CoV-2 Antigen Tests w... | Class II | WHPM Inc. |
| Mar 3, 2022 | DAVINCI | Energy SynchroSeal da Vinci X' da Vinci Xi', Single Use Instrument ... | Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection c... | Class II | Intuitive Surgical, Inc. |
| Mar 3, 2022 | VESSEL SEALER EXTEND, Single-Use Instrument EndoWrist Instruments 8mm, da Vin... | Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection c... | Class II | Intuitive Surgical, Inc. |
| Mar 1, 2022 | FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10 | The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter M... | Class II | Edwards Lifesciences, LLC |
| Mar 1, 2022 | FORE-SIGHT ELITE Absolute Tissue Oximeter, REF: 01-06-3000, Contents: Monito... | The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter M... | Class II | Edwards Lifesciences, LLC |
| Feb 28, 2022 | Philips Respironics V60 Ventilator (All Models, including: 1053613, 1053614, ... | Ventilator units have an issue impacting an internal electrical circuit within the ventilators, t... | Class I | Respironics California, LLC |
| Feb 28, 2022 | Philips Respironics V60 Plus Ventilator (All Models, including: 1138747, 1137... | Ventilator units have an issue impacting an internal electrical circuit within the ventilators, t... | Class I | Respironics California, LLC |
| Feb 28, 2022 | Philips Respironics V680 Ventilator (All Models, including: P/N 1103044 (8500... | Ventilator units have an issue impacting an internal electrical circuit within the ventilators, t... | Class I | Respironics California, LLC |
| Feb 22, 2022 | Illumina NextSeq 550Dx, REF: 20005715, CE, IVD | There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection... | Class II | Illumina, Inc. |
| Feb 21, 2022 | Silatrix Oral Gel Polymerized Sucralfate Gel 10% (1 gm/10 gm | Due to failure of microbial testing for total aerobic microbial count (TAMC). | Class II | SA3, LLC |
| Feb 3, 2022 | MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Nu... | Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or be... | Class III | Volcano Corp |
| Jan 31, 2022 | INFX-8000V. For radiographic and fluoroscopic studies and intervention. | The fluoroscopic dose rate might exceed the conforming value during biplane fluoroscopy. | Class II | Canon Medical System, USA, INC. |
| Jan 27, 2022 | MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741... | New or replacement insulin pumps are not pre-programmed with basal rates or other verified settin... | Class II | Medtronic MiniMed |
| Jan 27, 2022 | The product is a handheld ultraviolet-C germicidal wand and contains two, nar... | Excessive ultraviolet-C radiation | Class II | IN MY BATHROOM LLC |
| Jan 27, 2022 | Panther Fusion Extraction Reagent-X, REF PRD-04477, For use with Panther Fusi... | Failed stability specifications for reagent kits that may result in false negative test results. | Class II | Hologic, Inc |
| Jan 24, 2022 | IntraSight Mobile IVUS Systems, Model: 797415, Part numbers: 300002018101, 30... | Multi-Modality Touch Screen Modules (MM-TSM), which are bedside/secondary control units, used as ... | Class III | Volcano Corp |
| Jan 24, 2022 | Philips Respironics V60 Ventilator Part Number 1053617 | A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a br... | Class I | Respironics California, LLC |
| Jan 24, 2022 | Philips Respironics V60 Plus Ventilator Part Number 1138747 | A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a br... | Class I | Respironics California, LLC |
| Jan 20, 2022 | Outset Tablo Console Model Number: PN-0003000 Model Number: PN-0006000 | Due to a component in the hemodialysis console there is the possibility of heat-related damage th... | Class II | Outset Medical, Inc. |
| Jan 20, 2022 | QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL ... | Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting i... | Class II | Microgenics Corporation |
| Jan 19, 2022 | SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software versi... | If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedura... | Class II | Volcano Corporation |
| Jan 14, 2022 | CIRRUS HD-OCT, Models 5000 AngioPlex and 6000 AngioPlex | Optic nerve head angiography scan to be turned off due to its distribution without pre-market cle... | Class II | Carl Zeiss Meditec, Inc. |
| Jan 14, 2022 | 1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general l... | A software defect in the camera control unit (CCU) will cause the image on the monitor to flip up... | Class II | Stryker Corporation |
| Jan 13, 2022 | COVID-19 IgG/IGM Antibody Test For Detection of COVID-19 IgG and IgM antibod... | COVID-19 test kits (Antibody Rapid Test kit and Antigen Rapid Test Kit) are not authorized, clear... | Class I | Lusys Laboratories, Inc. |
| Jan 13, 2022 | LuSys 2nd Generation of COVID-19 Viral Antigen Test For all mutant variety Ca... | COVID-19 test kits (Antibody Rapid Test kit and Antigen Rapid Test Kit) are not authorized, clear... | Class I | Lusys Laboratories, Inc. |
| Jan 13, 2022 | Luscient Diagnostics 2nd Generation of COVID-19 Viral Antigen Test For all mu... | COVID-19 test kits (Antibody Rapid Test kit and Antigen Rapid Test Kit) are not authorized, clear... | Class I | Lusys Laboratories, Inc. |
| Jan 10, 2022 | MODEL: X-RAY R/F SYSTEM FLUOROspeed X1 | It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1... | Class II | Shimadzu Medical Systems |
| Jan 10, 2022 | MODEL: X-RAY TV SYSTEM SONIALVISION G4 | It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1... | Class II | Shimadzu Medical Systems |
| Jan 10, 2022 | MODEL: X-RAY TV SYSSTEM SONIALVISION safire17 | It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1... | Class II | Shimadzu Medical Systems |