FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10
FDA Device Recall #Z-0835-2022 — Class II — March 1, 2022
Recall Summary
| Recall Number | Z-0835-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | March 1, 2022 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Edwards Lifesciences, LLC |
| Location | Irvine, CA |
| Product Type | Devices |
| Quantity | 8389 units |
Product Description
FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10
Reason for Recall
The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, CANARY ISLANDS, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Northern Ireland, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates (UAE), United Kingdom, Vietnam.
Lot / Code Information
FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10, UDI: 00690103208573; All serial numbers.
Other Recalls from Edwards Lifesciences, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2085-2025 | Class I | Brand Name: FEM-FLEXII Product Name: Femoral A... | May 14, 2025 |
| Z-2080-2025 | Class I | Brand Name: Optisite Arterial Cannula Product ... | May 14, 2025 |
| Z-2084-2025 | Class I | Brand Name: FEM-FLEX II Product Name: Femoral ... | May 14, 2025 |
| Z-2081-2025 | Class I | Brand Name: Optisite Arterial Cannula Product ... | May 14, 2025 |
| Z-2083-2025 | Class I | Brand Name: FEM-FLEX II Product Name: Femoral ... | May 14, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.