Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q

FDA Device Recall #Z-0988-2022 — Class II — March 10, 2022

Recall Summary

Recall Number Z-0988-2022
Classification Class II — Moderate risk
Date Initiated March 10, 2022
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm St. Jude Medical, Cardiac Rhythm Management Division
Location Sylmar, CA
Product Type Devices
Quantity 1

Product Description

Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q

Reason for Recall

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.

Lot / Code Information

REF# CDDRA500Q SN# 111018237, Exp Date: 31-Oct-2022 Material# 600115496 UDI# 05415067032003

Other Recalls from St. Jude Medical, Cardiac Rhythm Mana...

Recall # Classification Product Date
Z-1496-2022 Class II Merlin 2 PCS MER37000 programmer Model MER3400 ... Jun 17, 2022
Z-1495-2022 Class II Merlin PCS 3650 programmer Model 3330 software Jun 17, 2022
Z-1497-2022 Class II Merlin.net model MN5000 Software Jun 17, 2022
Z-0990-2022 Class II Gallant HF Implantable Cardioverter Defibrillat... Mar 10, 2022
Z-0965-2022 Class II Merlin 2 PCS MER3700 programmer Model MER3400 s... Mar 10, 2022

Frequently Asked Questions

Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?