Digital clamp meters and multimeters

CPSC Recall #11-067 — December 13, 2010

Recall #11-067 Date: December 13, 2010 Remedy: Replace

Product

Digital clamp meters and multimeters

Description

This recall involves Extech digital clamp meters with model numbers EX612, EX613, EX622 and EX623 and multimeters with model numbers EX540, EX542 and EX570. "Extech" and the model number can be found on the front of the product and the serial number on the rear. Serial numbers that begin with "A" are not included in the recall.

Hazard

When the battery runs low, the meters can fail to give an accurate voltage reading, resulting in the operator falsely believing the electrical power is off or low. This poses an electrocution hazard.

Incidents / Injuries

Extech received one report of a meter displaying an incorrect voltage reading. No injuries have been reported.

Remedy

Consumers should immediately stop using the recalled digital clamp meters and multimeters and contact Extech for a free replacement meter.

Units Affected

5,100 units

Sold At

Industrial and electrical distributors and wholesalers nationwide and at online tool and test equipment retailers from January 2008 to November 2010 for between $150 and $300.

Manufacturer

Extech Instruments, of Waltham, Mass.

Manufactured In

China