Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
FDA Device Recall #Z-2558-2025 — Class II — July 8, 2025
Recall Summary
| Recall Number | Z-2558-2025 |
| Classification | Class II — Moderate risk |
| Date Initiated | July 8, 2025 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Boston Scientific Neuromodulation Corporation |
| Location | Valencia, CA |
| Product Type | Devices |
| Quantity | 25,260 |
Product Description
Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Reason for Recall
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.
Lot / Code Information
UDI-DI: 08714729985044, 08714729985051. Document Number/Revision: 92328632-09, Rev A
Other Recalls from Boston Scientific Neuromodulation Cor...
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2340-2025 | Class II | Burr Hole Cover Kit, REF: M365DB4600C0; Burr H... | Jul 8, 2025 |
| Z-2485-2025 | Class II | Superion IDS Kit Superion Indirect Decompressi... | Jun 6, 2025 |
| Z-2617-2024 | Class II | WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, R... | Jul 17, 2024 |
| Z-1890-2024 | Class II | Vercise Genus Deep Brain Stimulation (DBS) Impl... | Apr 18, 2024 |
| Z-1972-2023 | Class II | Driver Instrument, part of the Superion Indirec... | May 12, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.