Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with t...

FDA Device Recall #Z-1972-2023 — Class II — May 12, 2023

Recall Summary

Recall Number Z-1972-2023
Classification Class II — Moderate risk
Date Initiated May 12, 2023
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Boston Scientific Neuromodulation Corporation
Location Valencia, CA
Product Type Devices
Quantity 21,533

Product Description

Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual

Reason for Recall

Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.

Distribution Pattern

US Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND

Lot / Code Information

Driver UDI-DI: 00884662000574, UPN: 102-9800, All Lots. Superion Indirect Decompression System IFU (92479815-02), Superion IDS Kit IFU (92479820-02), Surgical Technique Manual (92479821-02)

Other Recalls from Boston Scientific Neuromodulation Cor...

Recall # Classification Product Date
Z-2558-2025 Class II Vercise Deep Brain Stimulation Systems Surgical... Jul 8, 2025
Z-2340-2025 Class II Burr Hole Cover Kit, REF: M365DB4600C0; Burr H... Jul 8, 2025
Z-2485-2025 Class II Superion IDS Kit Superion Indirect Decompressi... Jun 6, 2025
Z-2617-2024 Class II WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, R... Jul 17, 2024
Z-1890-2024 Class II Vercise Genus Deep Brain Stimulation (DBS) Impl... Apr 18, 2024

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.