C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon...

FDA Device Recall #Z-0970-2019 — Class II — December 3, 2018

Recall Summary

Recall Number Z-0970-2019
Classification Class II — Moderate risk
Date Initiated December 3, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm PENTAX of America Inc
Location Redwood City, CA
Product Type Devices
Quantity 45

Product Description

C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus.

Reason for Recall

The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur.

Distribution Pattern

Worldwide Distribution: US (nationwide) to states of:: CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, WA;; and countries: Canada, Germany, and Netherlands.

Lot / Code Information

All lot numbers

Other Recalls from PENTAX of America Inc

Recall # Classification Product Date
Z-2329-2025 Class II Pentax Medical Video Processor; Model Number: ... Jul 16, 2025
Z-1236-2025 Class II PENTAX Medical Video Colonoscope- Intended to ... Jan 29, 2025
Z-1238-2025 Class II Pentax Medical Video Processor- Intended to be ... Jan 29, 2025
Z-1237-2025 Class II PENTAX Medical Video Upper GI Scope-Intended to... Jan 29, 2025
Z-2449-2021 Class II Gastroscope Family # 1-Gastroscopes without a W... Jul 14, 2021

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.