Pentax Medical Video Processor; Model Number: EPK-i8020c;
FDA Recall #Z-2329-2025 — Class II — July 16, 2025
Product Description
Pentax Medical Video Processor; Model Number: EPK-i8020c;
Reason for Recall
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
Recalling Firm
Pentax of America Inc — Montvale, NJ
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
144 units
Distribution
US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.
Code Information
Model Number: EPK-i8020c; UDI-DI: 04961333247974; Serial Numbers: B0023Z1657, D0023Z0245, D0023Z0206, D0023Z0125, D0023Z0123, D0023Z0119, D0023Z0118, D0023Z0117, D0023Z0114, D0023Z0112, D0023Z0109, D0023Z0108, D0023Z0011, C0023Z0087, C0023Z0085, C0023Z0074, C0023Z0072, C0023Z0021, B0023Z1712, B0023Z1711, B0023Z1670, B0023Z1665, B0023Z1659, B0023Z1653, B0023Z1628, b0023z1501, B0023Z1195, B0023Z1707, C0023Z0084, B0023Z1732, B0023Z1669, D0023Z0137, D0023Z0141, D0158Z0049, D0158Z0033, E0023Z0022, D0158Z0053, D0158Z0019, E0023Z0026, E0023Z0008, D0023Z0248, D0023Z0241, D0023Z0222, D0023Z0155, B0023Z1276, C0023Z0082, D0023Z0233, B0023Z1461, D0023Z0139, D0023Z0140, D0023Z0122, B0023Z1709, E0023Z0020, E0023Z0009, E0023Z0002, D0023Z0142, E0023Z0024, E0023Z0018, E0023Z0013, E0023Z0004, C0023Z0095, C0023Z0078, E0023Z0007, E0158Z0057, E0158Z0035, B0023Z1735, B0023Z1717, B0023Z1648, B0023Z1639, B0023Z1631, C0158Z0047, C0158Z0042, B0023Z1481, B0023Z1478, B0023Z1451, B0023Z1335, B0023Z1193, B0023Z1288, B0023Z1258, B0023Z1296, B0023Z1290, B0023Z1286, B0023Z1280, B0023Z1279, B0023Z1256, B0023Z1250, B0023Z1245, B0023Z1676, B0023Z1573, B0023Z1565, B0023Z1656, B0023Z1694, D0158Z0055, B0023Z1577, B0023Z1570, B0023Z1682, B0023Z1678, B0023Z1661, B0023Z1637, D0158Z0048, D0158Z0047, E0023Z0016, E0023Z0001, D0023Z0071, D0023Z0150, D0023Z0148, D0023Z0143, B0023Z1277, D0023Z0152, D0023Z0144, D0023Z0106, D0023Z0102, B0023Z1667, B0023Z1652, B0023Z1645, B0023Z1727, B0023Z1723, B0023Z1702, E0023Z0003, D0023Z0244, D0023Z0232, C0023Z0091, D0023Z0208, B0023Z1655, D0023Z0219, D0023Z0017, D0023Z0008, C0023Z0098, C0023Z0097, C0023Z0096, C0023Z0094, C0023Z0092, B0023Z1716, B0023Z1698, B0023Z1697, B0023Z1696, B0023Z1671, B0023Z1255, B0023Z1247, B0023Z1291, B0023Z1289, B0023Z1287, B0023Z1285, B0023Z1243;
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated