Browse Device Recalls
530 FDA medical device safety recalls.
FDA Medical Device Recall Enforcement Database
Browse 530 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.
Medical Device Recalls
Browse 530 FDA device recalls.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Mar 25, 2016 | The Plum A+ is a dual-line volumetric infusion system designed to meet the gr... | Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were att... | Class II | Hospira Inc. |
| Feb 29, 2016 | Medica ISE Module Calibrant A; 500 mL - 140 mmol/L Na+, 125 mmol/L Cl-, buffe... | Lot was not functioning properly and not exhibiting the correct physical properties. | Class III | AMS Diagnostics, LLC |
| Feb 4, 2016 | Siemens Healthcare Diagnostic ADVIA Centaur Calibrator E- 6- Pack (Reference)... | ADVIA Centaur Systems Calibrator E Lot 42 and 43 Bias with the Testosterone Assay | Class II | Siemens Healthcare Diagnostics, Inc |
| Feb 4, 2016 | Siemens Healthcare Diagnostics ADVIA Centaur Calibrator E - 2-Pack Catalog... | ADVIA Centaur Systems Calibrator E Lot 42 and 43 Bias with the Testosterone Assay | Class II | Siemens Healthcare Diagnostics, Inc |
| Feb 4, 2016 | Siemens Healthcare Diagnostics ADVIA Centaur Calibrator E- 6- Pack Catalog Nu... | ADVIA Centaur Systems Calibrator E Lot 42 and 43 Bias with the Testosterone Assay | Class II | Siemens Healthcare Diagnostics, Inc |
| Jan 20, 2016 | LeadCare II Blood Lead Test Kit, containing LeadCare II Lead Controls Prod... | LeadCare II Test Kits controls out of range | Class II | Magellan Diagnostics, Inc. |
| Jan 11, 2016 | Product 72 consists of all product code: JDI and same usage: Item no: 805... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 24 consists of all product under product code: HRS and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 44 consists of all product under product code: HWC and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 23 consists of all product code: JDI and same usage: Item no: 8011... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 21 consists of all product under product code: JDI and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 68 consists of all product under product code: JDI and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 61 consists of all product under product code: JWH and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 2 consists of all products product code LPH, and same usage: 8018022... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 6 consists of all products under product code HWC, and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 55 consists of all product under product code: HSB and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 58 consists of all product under product code: HRS and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 38 consists of all product under product code: HSB and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 33 consists of all product under product code: JDI and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 11 consists of all product under product code JDI , and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 57 consists of all product code: JDI and same usage: Item no: 7634... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 39 consists of all product under product code: JDI and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 50 consists of all product under product code: JWH and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 9 consists of all product under HRS, and same usage: Item no: 4749... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 32 consists of all product under product code: HWC and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 41 consists of all product under product code: KTT and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 26 consists of all product under product code: JWH and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 49 consists of all product under product code: JWH and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 47 consists of all product under product code: LPH and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 40 consists of all product under product code: HSB and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 71 consists of all product under product code: JDI and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 13 consists of all product under product code HWC , and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 22 consists of all product under product code: JDI and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 20 consists of all prod oct under poduct code: KTT and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 66 consists of all product under product code: JDI and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 70 consists of all product under product code: JDI and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 63 consists of all product under product code: LPH and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 53 consists of all product under: product code: JWH and same usage... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 60 consists of all product under product code: HSB and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 30 consists of all product under product code: KTT and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 69 consists of all product under product code: JDI and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 46 consists of all product under product code: HWC and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 16 consists of all product under poduct code: KWY and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 4 consists of all products under product code HWC, and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 43 consists of all product under product code: HSB and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 42 consists of all product under product code: HSB and same usage: ... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 56 consists of all product under product code: HWC and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 15 consists of all product code JDI , and same usage: Item no: 80... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 3 consists of all products under product code JDI, and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
| Jan 11, 2016 | Product 8 consists of all products under product code HWC, and same usage: I... | As a result of the insufficient sealer calibration data for product packaged in firm's Building I... | Class II | Zimmer Manufacturing B.V. |
Frequently Asked Questions
When a device safety issue is identified, the manufacturer or FDA initiates a recall. The FDA publishes an enforcement report detailing the device, the reason for the recall, and distribution. Healthcare facilities and providers are notified to check their inventory and take corrective action.
All types — from simple devices like blood glucose test strips and surgical gloves to complex devices like pacemakers, defibrillators, insulin pumps, hip implants, and ventilators. Software-driven devices are increasingly common in recalls due to bugs and cybersecurity vulnerabilities.
Use the search and filter tools on this page. Each recall entry includes the device description, model numbers, and distribution details so you can determine if your device is affected. Contact your healthcare provider if you find a match.