Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certai...

FDA Device Recall #Z-2525-2016 — Class II — June 22, 2016

Recall Summary

Recall Number Z-2525-2016
Classification Class II — Moderate risk
Date Initiated June 22, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Healthcare Diagnostics Inc
Location Norwood, MA
Product Type Devices
Quantity 977 units

Product Description

Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diagnostic use. Assays include ~hCG, CRP, CKMB, cTNI, D-Dimer, Myoglobin, and pBNP. SMN: 10444834, 10453531

Reason for Recall

Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when a TestPak is due to expire within 48 hours.

Distribution Pattern

Worldwide Distribution -- USA, U.A.E., Algeria, Armenia, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Canada, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Libya, Lithuania, Malaysia, Mexico, Namibia, P.R. China, Poland, Portugal, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., and United Kingdom.

Lot / Code Information

All software versions and serial numbers

Other Recalls from Siemens Healthcare Diagnostics Inc

Recall # Classification Product Date
Z-2272-2025 Class II epoc BGEM BUN Test Card [25pk]. Material Number... Jul 7, 2025
Z-2230-2025 Class II RAPIDPoint 500 Systems Wash/Waste Cartridge (4 ... Jun 26, 2025
Z-1673-2025 Class II Stratus CS Acute Care cTNI TestPak.IVD test for... Mar 31, 2025
Z-1684-2025 Class II epoc BGEM BUN Test Card (25 pk) for the epoc Bl... Mar 19, 2025
Z-1685-2025 Class II epoc BGEM Crea Test Card (25 pk) for the epoc B... Mar 19, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.