Medical Device Recalls
Browse 3,419 FDA device recalls in MA.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Mar 16, 2018 | Intera 1.0T; 1) Stellar Model 781102, 2) Pular Model 781103 Product Usage... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | Ingenia 1.5T CX, Models 781261 & 781262 Product Usage: Indicated for use ... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | Multiva systems; 1) Magnets Multiva 16 Model 781487, 2) Magnets 8 Model 78148... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | Panorama 1.0T, Model 781250 Product Usage: Indicated for use as a diagnos... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | Intera 0.5T Standard, Model 781101 Product Usage: Indicated for use as a ... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | T10-NT, Model 78107 Product Usage: Indicated for use as a diagnostic devi... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | Achieva 1.5T: 1) Nova (Dual), Model 781173, 2) Achieva 1.5T Initial Systems M... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | Conversion SmarthPath to dStream for 1.5T, Model 781260 Product Usage: In... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | Multiva systems, 1) Multiva 1.5T 8 R5 Model 781072, 2) Multiva 1.5T 16 R5 Mod... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 16, 2018 | T5-NT, Model 78104 Product Usage: Indicated for use as a diagnostic devic... | Potential risk for helium gas inside the MR examination room during a magnet quench | Class II | Philips Electronics North America Corporation |
| Mar 14, 2018 | Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen m... | After continuous operation for more than one and a half days, the image on the large screen monit... | Class II | Philips Electronics North America Corporation |
| Mar 13, 2018 | Genesis Surgical Cassette Tapered Implants are intended for placement fo... | The surgical cassette contained extra-large white grommets making the 3.8 mm and 6.5mm dental bur... | Class II | Keystone Dental Inc |
| Mar 13, 2018 | Medtronic DxTerity Diagnostic Catheter, JL 4.0, REF DXT5JL40, marked with 6F ... | The color coded French size indicator on the outer box may incorrectly reflect a 6 French size, r... | Class II | Medtronic Vascular |
| Feb 26, 2018 | WET-FIELD OSHER THERMODOT MARKER, 25GA FINE TIP, CURVED; Part number 221270. ... | Incorrect tip was used during manufacturing of the Osher Thermodot Marker. The markers have a nee... | Class II | Beaver-Visitec International Inc. |
| Feb 26, 2018 | Wet-Field Hemostatic Eraser Bipolar 25G, fine tip, straight, 5/SP; Part Numbe... | Products were packaged with incorrect labeling, | Class II | Beaver-Visitec International Inc. |
| Feb 21, 2018 | Infinity Acute Care System (IACS) Monitoring Solution, Catalog Numbers: MS188... | The invasive pressure visual and audible alarms may not activate. | Class II | Draegar Medical Systems, Inc. |
| Feb 21, 2018 | ADVIA Centaur CA 125II Assay; Cat No. 09427226, SMN 10310443 (1-pack/100 test... | Upon dilution, some patient samples may exhibit over-recovery outside the representative data pro... | Class II | Siemens Healthcare Diagnostics, Inc |
| Feb 21, 2018 | F-T3SYSTEM, T3 (Trumpf Table Top) System The device is a surgical table ... | The recalling firm has become aware that the rail section clamps used to attach the leg support o... | Class II | Allen Medical Systems |
| Feb 12, 2018 | Capio RP; Model #M0068321010. Intended for use in general suturing applica... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | Capio SLIM Suture Capturing Device, Model #M0068318250. Intended for use i... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | Capio Standard Box 4 (Capio PC); Model #M0068312321. Intended for use in g... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | Posterior LITE w/ Capio SLIM; Model #M0068318150. Indicated for tissue rei... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | Uphold LITE w/ Capio SLIM; Model #M0068318170. Indicated for tissue reinfo... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | Uphold Vaginal Support System; Model #M0068317080. Indicated for tissue re... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | PFR Kit- Pinnacle, Posterior; Model #M0068317100. Indicated for tissue rei... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | Capio CL; Model #M0068311350. Indicated for the placement of suture in a v... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | PFR Kit- Pinnacle, Anterior Apical; Model #M0068317100. Indicated for tiss... | Class II | Boston Scientific Corporation | |
| Feb 12, 2018 | Capio Open Access Box 4 (Capio OPC); Model #M0068311251. Intended for use ... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 12, 2018 | Capio SLIM Box 5, Model #M0068318261; Intended for use in general suturin... | Potential for Capio sutures to break and /or detach. | Class II | Boston Scientific Corporation |
| Feb 7, 2018 | GloCyte RBC Reagent - Product Usage: The GloCyte Automated Cell Counter Syste... | RBC Reagent failure. High fluorescent counts during RBC Reagent + Diluent QC check that will resu... | Class II | Advanced Instruments, LLC |
| Feb 7, 2018 | HeartStart XL+ Defibrillator/Monitor The HeartStart XL+ is intended for us... | Update XL¿ device software to version A.03. This includes enhancements to the Operational Check... | Class III | Philips Electronics North America Corporation |
| Jan 29, 2018 | NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M5... | May contain endotoxin levels which have been confirmed to exceed the ANSI/AAMI quality standard d... | Class II | NxStage Medical, Inc. |
| Jan 29, 2018 | Philips QCPR Meter - used with MRx Product Usage: The Philips CPR meter i... | Inform users of correct placement of the disposable adhesive pad on the QCPR meter and inform use... | Class II | Philips Electronics North America Corporation |
| Jan 29, 2018 | Philips QCPR Meter - used with FR3 Product Usage: The Philips CPR meter i... | Inform users of correct placement of the disposable adhesive pad on the QCPR meter and inform use... | Class II | Philips Electronics North America Corporation |
| Jan 25, 2018 | O-arm 1000 mobile image-intensified fluoroscopic x-ray system; (1) Model: BAS... | Software update ("SW Update 3.2.1") is being implemented to address the following issues: Inabili... | Class II | Medtronic Navigation, Inc.-Littleton |
| Jan 24, 2018 | Liberty Select Cycler (SW v.2.8.7), Material Number RTLR108343 Product Usa... | The recalling firm identified a software issue related to the Patient Line Check (PLC) which may ... | Class II | Fresenius Medical Care Renal Therapies Group, LLC |
| Jan 24, 2018 | Stage-1 Tissue Punch, Disposable, 4.5mm; catalog number: ISS141K. Intended... | The cutting edge of tissue punches may be potentially damaged with small dents and nicks. | Class II | Keystone Dental Inc |
| Jan 12, 2018 | Solid state x-ray imager (flat panel/digital imager) Solid state x-ray i... | While transporting the portable machine with the retrofit equipment in a backwards motion, there ... | Class II | NeuroLogica Corporation |
| Jan 9, 2018 | LeadCare Plus Blood Lead Analyzer Part Number: 82-0001 Intended for the q... | The incorrect placement of AA batteries into certain LeadCare Plus Blood Lead Analyzer may cause ... | Class II | Magellan Diagnostics, Inc. |
| Jan 9, 2018 | LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P,... | The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause th... | Class II | Magellan Diagnostics, Inc. |
| Jan 4, 2018 | Allura Xper systems with Certeray X-ray Generator located in regions with hig... | There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these produc... | Class II | Philips Electronics North America Corporation |
| Jan 3, 2018 | Intellivue iX Information Center Software - All PIIC iX Surveillance stations... | Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable ... | Class II | Philips Electronics North America Corporation |
| Dec 22, 2017 | LATARJET EXPERIENCE Sharp Curved Ostetome Product Usage: The Latarjet EXP... | Reports of product loosening or coming unscrewed from the handle. | Class II | DePuy Mitek, Inc., a Johnson & Johnson Co. |
| Dec 20, 2017 | Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: ... | Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), mi... | Class II | Philips Electronics North America Corporation |
| Dec 20, 2017 | SpyScope DS Access and Delivery diagnostic biliary catheter, M00546600 | Certain lots of devices may have the working channel sleeve protruding from the camera cap which ... | Class II | Boston Scientific Corporation |
| Dec 18, 2017 | Terumo Cardiovascular Procedure Kits containing Pall LG6NS LeukoGuard¿ Leukoc... | Possible blood leaks through the hydrophobic portion of the Pall LG6NS LeukoGuard Leukocyte Reduc... | Class II | Terumo Cardiovascular Systems Corp |
| Dec 15, 2017 | Confidence Plus Kit Spinal Cement System, Product Codes: 283910000 (GTIN: (01... | Certain lots of the spinal cement may contain incorrect amounts of contrast medium so that the pr... | Class II | DePuy Spine, Inc. |
| Dec 14, 2017 | 2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodial... | When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service ... | Class II | Fresenius Medical Care Renal Therapies Group, LLC |
| Dec 12, 2017 | Malecot Nephrostomy Catheter Set, Product Number: M0064101010, GTIN: 08714729... | Certain Malecot Nephrostomy Catheters are breaking at the mid-shaft bond during use. | Class II | Boston Scientific Corporation |
| Dec 12, 2017 | Re-Entry" Malecot Nephrostomy Catheter Set, Product Number: M0064101040, GTIN... | Certain Malecot Nephrostomy Catheters are breaking at the mid-shaft bond during use. | Class II | Boston Scientific Corporation |