Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventio...

FDA Device Recall #Z-1589-2018 — Class II — March 14, 2018

Recall Summary

Recall Number Z-1589-2018
Classification Class II — Moderate risk
Date Initiated March 14, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Philips Electronics North America Corporation
Location Andover, MA
Product Type Devices
Quantity 682

Product Description

Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray system Product Usage: The Allura Xper ED series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, eg peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placement, embolization and thrombolysis. Cardiac imaging applications including diagnostics, interventional minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper ED series is compatible with a hybrid operating room.

Reason for Recall

After continuous operation for more than one and a half days, the image on the large screen monitor may freeze for approximately 15 seconds after which the system will restore itself.

Distribution Pattern

Worldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The product was distributed to the following countries: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Croatia, Cuba, Czech Republic, Denmark, Estonia, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Republic of Korea, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, Norway, Pakistan, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, and United Kingdom.

Lot / Code Information

Product name: Allura Xper R8.1 .25.0 Allura Xper R8.1 .25.1 Allura Xper R8.1 .25.5 Allura Xper R8.2.25.0 Allura Xper R8.2.25.5 Allura Xper R8.2.27 UNIQ Ri .0.10 UNIQ Ri .0.10.5 Product code: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 7322039, 722058, 722059.

Other Recalls from Philips Electronics North America Co...

Recall # Classification Product Date
Z-0614-2019 Class II Xper Flex Cardio Physiomonitoring System, Softw... Nov 15, 2018
Z-0592-2019 Class II IntelliVue MX40 Patient Monitor, Model Nos. 865... Oct 10, 2018
Z-3227-2018 Class II Allura Xper F010, System Code 722003 Vascul... Jul 16, 2018
Z-3218-2018 Class II INTEGRIS SUITE, System Code 722199 Vascular... Jul 16, 2018
Z-3225-2018 Class II Allura Xper FD1O C, System Code 722001 Vasc... Jul 16, 2018

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.