| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Mar 1, 2022 | AccessPLUS Large Bore Hemostasis Valves, Catalog Code: 580758001, 580774001, ... | A design change made to a hemostasis valve to improve manufacturability caused an internal gap be... | Class II | Merit Medical Systems, Inc. |
| Feb 17, 2022 | Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a co... | Reports have been received that after normal or routine lateral forces are applied to the head fi... | Class II | Deerfield Imaging, Inc. |
| Feb 10, 2022 | Nextra Hammertoe Correction System 4.5 Middle and 3.2 Proximal Saw Cut SKU:... | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect ... | Class II | Nextremity Solutions |
| Feb 10, 2022 | Nextra Hammertoe Correction System 4.5 Middle 3.2 Proximal Kit SKU: NX-4532... | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect ... | Class II | Nextremity Solutions |
| Feb 10, 2022 | Nextra Hammertoe Correction System 3.5 Middle 3.2 Proximal Kit SKU: NX-3532... | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect ... | Class II | Nextremity Solutions |
| Feb 10, 2022 | Nextra Hammertoe Correction System Driver SKU: NX-DR Orthopedic surgical ... | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect ... | Class II | Nextremity Solutions |
| Feb 10, 2022 | Nextra Hammertoe Correction System 3.5 Middle and 3.2 Proximal Saw Cut SKU:... | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect ... | Class II | Nextremity Solutions |
| Jan 27, 2022 | HeartWare Ventricular Assist Device (HVAD) System | Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and B... | Class II | Heartware, Inc. |
| Dec 17, 2021 | Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon C... | The firm has received six (6) complaint reports from customers indicating difficulty removing the... | Class II | Covidien Llc |
| Dec 6, 2021 | LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number L... | Customers reported that the device jaws were difficult to open or would not open following applic... | Class II | Covidien Llc |
| Nov 22, 2021 | WIRION EMBOLIC PROTECTION SYSTEM, Delivery Catheter and Filter and Retrieval ... | Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under s... | Class I | Cardiovascular Systems Inc |
| Nov 19, 2021 | Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set... | Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest w... | Class II | Medtronic Sofamor Danek USA Inc |
| Nov 10, 2021 | SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825 | Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B... | Class II | Beckman Coulter Inc. |
| Nov 9, 2021 | Torosa Saline-Filled Testicular Prosthesis (Size Small) | Product is mislabeled for size. | Class II | Coloplast Manufacturing US, LLC |
| Nov 4, 2021 | Puritan Bennett 980 Series Ventilator, 980S1ENDICU | Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, w... | Class I | Covidien Llc |
| Nov 4, 2021 | Puritan Bennett 980 Series Ventilator, 980S3ENDICU | Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, w... | Class I | Covidien Llc |
| Oct 14, 2021 | Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0 | When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, ... | Class II | FujiFilm Healthcare Americas Corporation |
| Oct 11, 2021 | BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 U... | Due to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results... | Class II | BioFire Diagnostics, LLC |
| Sep 21, 2021 | ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoline... | Due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in ... | Class II | Olympus Corporation of the Americas |
| Sep 13, 2021 | Medtronic Bio-Medicus Insertion Kits, Model number 96553, Pediatric Insertion... | There was a label mix-up between two model numbers resulting in the incorrect guidewire being con... | Class II | Medtronic Perfusion Systems |
| Aug 31, 2021 | Medtronic sterile single-use blades that are components of the StraightShot M... | There was an increase in complaints related to tip breaks and wobble/vibration. | Class II | Medtronic Xomed, Inc. |
| Aug 31, 2021 | Medtronic sterile single-use blades that are components of the StraightShot M... | There was an increase in complaints related to tip breaks and wobble/vibration. | Class II | Medtronic Xomed, Inc. |
| Aug 24, 2021 | McKesson PREVENT HT Safety Blood Collection Needles, 21G 1 1/4 inch, MFR # 16... | Potential for difficulty in safety mechanism activation. | Class II | Zhejiang Kindly Medical Devices Co., Ltd. |
| Aug 24, 2021 | McKesson PREVENT HT Safety Blood Collection Needles, 22G 1 1/4 inch, MFR # 16... | Potential for difficulty in safety mechanism activation. | Class II | Zhejiang Kindly Medical Devices Co., Ltd. |
| Aug 24, 2021 | McKesson PREVENT HT Safety Winged Blood Collection Needle with Tube Holder, M... | Potential for difficulty in safety mechanism activation. | Class II | Zhejiang Kindly Medical Devices Co., Ltd. |
| Aug 24, 2021 | HENRY SCHEIN Safety Blood Collection Needle 22Gx 1 1/4 Inch, REF 570-2068, 10... | Potential for difficulty in safety mechanism activation. | Class II | Zhejiang Kindly Medical Devices Co., Ltd. |
| Aug 24, 2021 | McKesson PREVENT HT Safety Winged Blood Collection Needle with Tube Holder, M... | Potential for difficulty in safety mechanism activation. | Class II | Zhejiang Kindly Medical Devices Co., Ltd. |
| Aug 24, 2021 | HENRY SCHEIN Safety Blood Collection Needle 21Gx 1 1/4 Inch, REF 570-2067, 10... | Potential for difficulty in safety mechanism activation. | Class II | Zhejiang Kindly Medical Devices Co., Ltd. |
| Aug 6, 2021 | Medtronic HVAD Pump Implant Kits | There is more force required to slide the driveline cover (boot cover) away from the connector af... | Class II | Heartware, Inc. |
| Aug 2, 2021 | BD ADD-ON Burette Set V/NV Needle-Free Valve, REF: 82113E-0006 | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD Extension Set SMALLBORE Tubing SmartSite VALVE, REF: 10010511; BD SmartSi... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD Alaris Pump Infusion Buerette Set Smallbore Tubing Smartsite Port (Burette... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD SmartSite Bag Access Device Needle-Free Valve, REF: 2300E; BD Bag access D... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD Gravity Burette Set 60 DP 3 SmartSite VALVES BALL VALVE DRIP CHAMBER, REF:... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD Syringe Set 20 DP MIC TBG 0.2MF CKV Anti-Siphon SmartSite VLV Sensing Disc... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD SmartSite, REF: 2000E CHINA; BD SmartSite Connector, REF: 2000E-04; BD Sma... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD BURETTE GRAVITY SET 60 DROP 2 Smartsite VALVES VENTED/NONVENTED, REF: 42163E | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD SmartSite 'Y' Extension Set 2 Needle-Free Valves, REF: 20019E7D | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD SmartSite Bag Access Device, REF: 2300E-0500; BD BAG ACCESS DEVICE CHECK V... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD Extension Set SMALLBORE Tubing with 2 SmartSite VALVES, REF: 20019E | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD Alaris Pump Infusion Buerette Set 0.2 Micron Filter Smallbore Tubing Smart... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD SE Burette Set 20 Drops 2 Injection Ports Vented/NonVented, REF: 72103E-0006 | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD Gravity IV Set 3-port Closed Manifold Back Check Valve 3 Needle-Free Y-sit... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Aug 2, 2021 | BD Gravity Burette Set 60 DROP 1.2 MF 2 SmartSite VALVES BALL VALVE DRIP CHAM... | Needle-free connector valves, either standalone or attached to extension sets, may have issues in... | Class II | CAREFUSION |
| Jul 14, 2021 | FLASH Ostial System, Dual Balloon Angioplasty Catheter, REF: OAB6014BA | Angioplasty system has a manufacturing issue that has the potential to result in difficulty defla... | Class II | Ostial Corporation |
| Jun 3, 2021 | HeartWare HVAD Implant Kit, REF MCS1705PU | Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of obser... | Class I | Heartware, Inc. |
| Jun 3, 2021 | HeartWare HVAD Driveline Extension Cable, REF 100US | Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of obser... | Class I | Heartware, Inc. |
| Jun 3, 2021 | HeartWare HVAD Implant Kit, REF 1104JP | Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of obser... | Class I | Heartware, Inc. |
| Jun 3, 2021 | HeartWare HVAD Pump Implant Kit, REF 1104 | Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of obser... | Class I | Heartware, Inc. |
| Jun 3, 2021 | HeartWare HVAD Pump Implant Kit, REF 1103 | Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of obser... | Class I | Heartware, Inc. |