Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIV...
FDA Device Recall #Z-1681-2022 — Class II — June 27, 2022
Recall Summary
| Recall Number | Z-1681-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 27, 2022 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | ICU Medical, Inc. |
| Location | San Clemente, CA |
| Product Type | Devices |
| Quantity | 454,635 |
Product Description
Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX 1.5ml,ChemoLock BAG SPIKE WITH ADDITIVE PORT/CL-13; ChemoLock Port/CL2100; ChemoLock Port, 5 Units/CL2100-5; ChemoLock Port/CL2100T; Clave Bag Spike w ChemoLock Port/CL2150; 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger/CL3011; 9 IN(23cm)APPX 2.7ml,EXT SET,ChemoLock PORT,2 CLAMPS,2 GRADUATED CONN/CL3364; Syringe Transfer Set w Clave, ChemoLock Port/CL-34; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511T; 30" (76cm) Appx 6.3 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter, ChemoLock w Red Cap, Hanger/CL3528; 7" (18 cm) Appx .93 ml, Ext Set w/ChemoLock, Y-Connector, Rotating Luer/CL3535T; Oncology Kit w/5" (13 cm) Add-On Set w ChemoLock Additive Port, Vented Cap, ChemoLock, Spinning Spiros w Red Cap/CL3538; Graduated Connector w ChemoLock Port/CL3900; ONCOLOGY KIT, EXT,ChemoLock PORT, GRAD CONN,TRANSFER,ChemoLock,VIAL SPIKE,ChemoLock PORT/CL3927; Oncology Kit w 5" (13 cm) Add-On Set w ChemoLock w Red Cap, Vented Cap, ChemoLock Port w Bag Spike, Spiros w Red Cap/CL3946; Oncology Kit w 17" (43 cm) Bifuse Ext Set w/ChemoLock w Red Cap, ChemoLock Port, ChemoLock Vented Vial Spike, 13mm ChemoLock/CL3947; Oncology Kit w ChemoLock Port Closed Vial Spike, ChemoLock, ChemoLock Port Bag Spike/CL3952; APPX 0.5ml, ChemoLock BAG SPIKE WITH ADDITIVE PORT,ChemoLock Port/CL3955;
Reason for Recall
Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.
Lot / Code Information
Item/UDI-DI/Lots: CL-10-10/10887709071022/5872839, 5891255; CL-12/10887709056975/5819981, 5856055, 5873186; CL-13/10887709056982/5850811; CL2100/5797867, 5816334, 5829882,5850510; CL2100-5/10887709122960/5754377; CL2100T/10887709099293/5816341, 5842229, 5850517; CL2150/10887709094670/5829900, 5850507; CL3011/10887709057019/5829885, 5842227; CL3364/10887709073408/5841261, 5850805; CL-34/10887709069777/5754437, 5829875; CL3511/10887709057026/5842231, 5861991, 5894477; CL3511T/10887709120010/5829831, 5891263; CL3528/10887709057040/5850500, 5872833, 5883793, 5894481, 5909807; CL3535T/10887709099613/5685350; CL3538/10887709071909/5935851; CL3900/10887709071411/5772697; CL3927/10887709075440/5842211; CL3946/10887709080291/5829895, 5861983; CL3947/10887709080307/5782844; CL3952/10887709081571/5782863,5816379; CL3955/10887709082509/5841209, 5850839; CL3960/10887709083278/5816328, 5842216; CL3963/10887709083827/5842214,5851775; CL4114/10887709091136/5754773, 5782864, 5816356; CL4130/10887709095530/5850515; CL4131/10887709095776/5895940; CL4136/10887709096131/5816325, 5850661; CL4143/10887709097213/5782848, 5873243, 5891266; CL4146-5/10887709097466/5872843, 5883800, 5909812; CL4153/10887709098722/5816348; CL4159/10887709121369/5881696; CL4179/10887709123578/5872687; CL-62/10887709057088/5816320,5825248; CL-70/10887709057095/5842228; CL-70-10/10887709070988/5816343, 5851774; CL-70-10T/10887709099477/5816353; CL-70S/10887709067155/5842225,5850512; CL-70S-5/10887709070995/5816326, 5851776; CL-70T/10887709099507/5816350; CL-72/10887709062440/5829832, 5849740, 5816312; CL-72T/10887709099552/5841181, 5849748; CL-76/10887709122878/5816323, 5842218; CL-80/10887709057125/5804034, 5842200, 5850499; CL-80S/10887709069593/5808063, 5816306, 5829020, 5829023, 5841982, 5845354, 5862839; CL-80S-10/10840619080577/5804045, 5816315; CL-80S-10T/10887709120539/5804051, 5816310, 5829022, 5841994, 5849630, 5850497, 5871336; CL-80S-5/10887709077116/5667054, 5829021, 5829024, 5841980, 5849474, 5849631, 5849632, 5849713, 5849714, 5849715, 5849716, 5849474; CL-80SL/10887709069630/5816317; CL-82/10887709085692/5804065; CLH-12/10887709080413/5842212
Other Recalls from ICU Medical, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1111-2026 | Class II | Tego Connector: REF: 011-D1000, 01C-D1000, 055-... | Dec 19, 2025 |
| Z-1094-2026 | Class II | Plum Duo Infusion System, List Number 40002-04-01 | Dec 15, 2025 |
| Z-1119-2026 | Class II | ICU Medical Plum Duo Precision IV Pump, List Nu... | Dec 15, 2025 |
| Z-1120-2026 | Class II | ICU Medical Plum Solo Precision IV Pump, List N... | Dec 15, 2025 |
| Z-1118-2026 | Class II | ICU Medical Plum Duo Infusion Pump, List Number... | Dec 15, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.