Browse Device Recalls

35,307 FDA medical device safety recalls.

FDA Medical Device Recall Enforcement Database

Browse 35,307 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.

Medical Device Recalls

Browse 35,307 FDA device recalls, Class II.

Clear
DateProductReasonClassFirm
Oct 16, 2023 PCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPE Specific devices may have an improperly repaired light guide connector due to missing adhesive to... Class II Aizu Olympus Co., Ltd.
Oct 16, 2023 CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE Specific devices may have an improperly repaired light guide connector due to missing adhesive to... Class II Aizu Olympus Co., Ltd.
Oct 16, 2023 PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE Specific devices may have an improperly repaired light guide connector due to missing adhesive to... Class II Aizu Olympus Co., Ltd.
Oct 13, 2023 Unity Total Knee System, Model Number 112.001.34. Primary and revision t... Potential for Unity CR Inserts Right size 6 from batch 529803 to be incorrectly labelled as Unity... Class II Corin Ltd
Oct 13, 2023 Unity CR Femur Right, Size 6, Model Number 112.001.32 Primary and revisi... Potential for Unity CR Inserts Right size 6 from batch 529803 to be incorrectly labelled as Unity... Class II Corin Ltd
Oct 13, 2023 AVANOS BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI, Product cod... Certain lots of BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI were labeled as not bei... Class II Avanos Medical, Inc.
Oct 13, 2023 with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 0060... Produced with unintended open slits on the side of the mouthpiece body, may result in insufficien... Class II Breas Medical, Inc.
Oct 12, 2023 Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM01... The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, on... Class II Medtronic Inc
Oct 12, 2023 Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS01... The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, on... Class II Medtronic Inc
Oct 12, 2023 Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbe... It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail ... Class II Maquet Medical Systems USA
Oct 12, 2023 Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM01... The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, on... Class II Medtronic Inc
Oct 12, 2023 Change Healthcare Radiology Solutions, software versions 14.0, 14.1, and 14.2... Change Healthcare has identified a software issue where Change Healthcare Radiology Solutions 14.... Class II CHANGE HEALTHCARE CANADA COMPANY
Oct 11, 2023 Curaplex manual resuscitators with integrated manometer, multiple accessory c... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 BD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump In... Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DE... Class II Becton Dickinson & Company
Oct 11, 2023 RANDOX Total Bilirubin. Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with To... Class II Randox Laboratories Ltd.
Oct 11, 2023 Ventlab, LLC RescuMed manual resuscitators with integrated manometer, multipl... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal l... May have increased water path restriction within the heat exchanger, leading to degraded heat exc... Class II Mc3 Inc
Oct 11, 2023 Ventlab, LLC BreathTech manual resuscitators with integrated manometer, multi... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 Ventlab, LLC STAT-Check manual resuscitators with integrated manometer, mult... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 Ventlab, LLC SafeSpot manual resuscitators with integrated manometer, multipl... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 Ventlab, LLC Horizon manual resuscitators with integrated manometer, multiple... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 Medline manual resuscitators with integrated manometer, multiple accessory co... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 Ventlab, LLC AirFlow manual resuscitator devices with integrated manometer, m... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 BD gravity sets (Product Name, Catalog #) 32-IN GRAV SET W/0.2 MICRON FILT... Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DE... Class II Becton Dickinson & Company
Oct 11, 2023 Ventlab, LLC V-Care manual resuscitators with integrated manometer, multiple ... A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreath... Class II SunMed Holdings, LLC
Oct 11, 2023 NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal lif... May have increased water path restriction within the heat exchanger, leading to degraded heat exc... Class II Mc3 Inc
Oct 11, 2023 BD Extension Sets (Product Name, Catalog #) BD SmartSite Filter Extension ... Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DE... Class II Becton Dickinson & Company
Oct 10, 2023 VITROS XT 7600 Integrated System Product Code 6844461 Running Software Versio... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 10, 2023 Automated Impella Controller (AIC), product number 0042-040-US with software ... Pump not detected as connected to controller due to software issue. Class II Abiomed, Inc.
Oct 10, 2023 VITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Co... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 10, 2023 VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Prod... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 10, 2023 Automated Impella Controller (AIC), product number 0042-000-US with software ... Pump not detected as connected to controller due to software issue. Class II Abiomed, Inc.
Oct 10, 2023 VITROS XT 3400 Chemistry System Product Code 6844458 Running Software Version... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 10, 2023 Stay Safe Cap Updating the labeling to include additional warnings and contraindications for potential side eff... Class II Fresenius Medical Care Holdings, Inc.
Oct 10, 2023 Xstar Kojo Safety Slit Knife, 2.5mm 45¿ Bevel Up, Part Number 373025 One lot of Part Number 373025, Xstar¿ Kojo Safety Slit Knife 2.5mm 45¿ bevel up, incorrectly cont... Class II Beaver Visitec International, Inc.
Oct 10, 2023 VITROS 4600 Chemistry System Product Code 6802445 and Refurbished Product Cod... During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence ... Class II Ortho-Clinical Diagnostics, Inc.
Oct 9, 2023 Compress Device Short Anchor Plug, 26 MM-For Correction of revision of unsucc... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Compress Device Short Anchor Plug, 24 MM-For Correction of revision of unsucc... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Compress Device Short Anchor Plug, 14 MM-For Correction of revision of unsucc... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Spectral CT on Rails, Software Version 5.1.0, Model Number 728334 This recall is part of a 2-issue recall for the same system. When using the IVC needle position b... Class II Philips North America
Oct 9, 2023 Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT... For oncology users: If the user performed off-set reconstruction on CT device a shift could be ob... Class II Philips North America Llc
Oct 9, 2023 Compress Device Short Anchor Plug, 28 MM-For Correction of revision of unsucc... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Compress Device Segmental Anchor Plug, 18 MM-For Correction of revision of un... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Compress Device Segmental Anchor Plug, 20 MM-For Correction of revision of un... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Compress Device Short Anchor Plug, 12 MM-For Correction of revision of unsucc... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Big Bore RT with software version V4.8.0.10421- Model: 728242 CT scanner ... For oncology users: If the user performed off-set reconstruction on CT device a shift could be ob... Class II Philips North America Llc
Oct 9, 2023 Compress Device Segmental Anchor Plug, 14 MM-For Correction of revision of un... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Philips Affiniti 70, Diagnostic Ultrasound System Model # 795210 Their is the potential that diagnostic ultrasound systems operating with software version 10.0 ma... Class II Philips Ultrasound, Inc.
Oct 9, 2023 Compress Device Segmental Anchor Plug, 10 MM-For Correction of revision of un... Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the d... Class II Biomet, Inc.
Oct 9, 2023 Spectral CT on Rails, Software Version 5.1.0, Model Number 728334 When preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the ope... Class II Philips North America

Frequently Asked Questions

When a device safety issue is identified, the manufacturer or FDA initiates a recall. The FDA publishes an enforcement report detailing the device, the reason for the recall, and distribution. Healthcare facilities and providers are notified to check their inventory and take corrective action.

All types, from simple devices like blood glucose test strips and surgical gloves to complex devices like pacemakers, defibrillators, insulin pumps, hip implants, and ventilators. Software-driven devices are increasingly common in recalls due to bugs and cybersecurity vulnerabilities.

Use the search and filter tools on this page. Each recall entry includes the device description, model numbers, and distribution details so you can determine if your device is affected. Contact your healthcare provider if you find a match.