| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Jun 17, 2019 | CS100 Intra-Aortic Balloon Pump, English, 110V; 0998-00-3013-53 For cardiova... | If battery maintenance is not performed appropriately, the battery may provide less than the mini... | Class I | Datascope Corp. |
| Jun 17, 2019 | CS300 Intra-Aortic Balloon Pump, English, 220V; 0998-00-3023-55 For cardiova... | If battery maintenance is not performed per instructions, the battery may provide less than the m... | Class I | Datascope Corp. |
| Jun 17, 2019 | Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), 0998-00-0800-53 For card... | Battery short-run times and unanticipated stoppage may be due to user not performing battery main... | Class I | Datascope Corp. |
| Jun 17, 2019 | CS300 Intra-Aortic Balloon Pump, English, 110V; 0998-00-3023-53 For cardiova... | If battery maintenance is not performed per instructions, the battery may provide less than the m... | Class I | Datascope Corp. |
| Jun 17, 2019 | Cardiosave Rescue IABP, 0998-00-0800-83 Intra-Aortic Balloon Pump For card... | If battery maintenance is not performed appropriately, the battery may provide less than the mini... | Class I | Datascope Corp. |
| Jun 14, 2019 | Babi.Plus 12.5 cm H2O Pressure Relief Manifold Catalog # 2691 for use with v... | Device vented gas below the stated pressure. | Class I | TELEFLEX-MORRISVILLE |
| Jun 7, 2019 | Sheridan Endotracheal Tube contained inside Centurion kit code TC7855. | Centurion is conducting a sub-recall for Teleflex Sheridan branded Endotracheal Tracheal Tubes du... | Class I | Centurion Medical Products Corporation |
| Jun 5, 2019 | Filter-Tips, 1500 ul (1024 tips/kitbox), REF 997024- REF: 997024 IVD: Used ... | Kit lots contain a small percentage of damaged Filter-Tips that are leaking when being used with ... | Class I | Qiagen Sciences LLC |
| May 30, 2019 | Babi.Plus Bubble PAP Valve with 12.5 cm H2O Pressure Relief Manifold, REF 100... | Reports of pressure relief manifold venting gas below the standard pressure. | Class I | Galemed Corporation |
| May 30, 2019 | Babi.Plus 12.5 cm H2O Pressure Relief Manifold, REF 1030, RX Only, UDI: 04710... | Reports of pressure relief manifold venting gas below the standard pressure. | Class I | Galemed Corporation |
| May 30, 2019 | Babi.Plus 12.5 cm H2O Pressure Relief Manifold with Gas Supply Tube and Oxyge... | Reports of pressure relief manifold venting gas below the standard pressure. | Class I | Galemed Corporation |
| May 29, 2019 | Quantum Pump Console | Potential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut d... | Class I | Spectrum Medical Ltd. |
| May 24, 2019 | Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5... | This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product... | Class I | Teleflex Medical |
| May 24, 2019 | Hudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6... | This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product... | Class I | Teleflex Medical |
| May 24, 2019 | Hudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Pro... | This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product... | Class I | Teleflex Medical |
| May 24, 2019 | Sheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Trachea... | This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product... | Class I | Teleflex Medical |
| May 24, 2019 | Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tub... | This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product... | Class I | Teleflex Medical |
| May 24, 2019 | Sheridan/HVT: 1) Novaplus 7.0 mm, Product Code V5-10314 2) Novaplus 7.5 mm,... | This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product... | Class I | Teleflex Medical |
| May 24, 2019 | Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/135cm, Ca... | Multiple complaints were received for balloons bursting below the rated burst pressure. Potentia... | Class I | Cook Inc. |
| May 24, 2019 | Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/50cm, Cat... | Multiple complaints were received for balloons bursting below the rated burst pressure. Potentia... | Class I | Cook Inc. |
| May 24, 2019 | Hudson RCI Sheridan: 1) Preformed 6.0 mm, Product Codes: a) 5-22212, b)... | This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product... | Class I | Teleflex Medical |
| May 24, 2019 | Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/80cm, Cat... | Multiple complaints were received for balloons bursting below the rated burst pressure. Potentia... | Class I | Cook Inc. |
| May 24, 2019 | Hudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6... | This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product... | Class I | Teleflex Medical |
| May 14, 2019 | IntraClude Intra-Aortic Occlusion Device, REF: ICF100, 10.5 Fr (3.5 mm) X 39.... | Possibility of balloon rupture in intra-aortic occlusion devices. If the balloon bursts during c... | Class I | Edwards Lifesciences, LLC |
| May 10, 2019 | Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code ... | This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to... | Class I | Medline Industries Inc |
| May 10, 2019 | Teleflex HUDSON RDI NEONATAL ConchaSmart Breathing Circuit with Dual Heated L... | Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonata... | Class I | Teleflex Medical |
| May 6, 2019 | Alaris Pump Infusion Set, Model Codes 10010453, 10010454, 10012144, 10015294,... | An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment... | Class I | Becton Dickinson & Company |
| May 3, 2019 | Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warm... | The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum leve... | Class I | Medline Industries Inc |
| May 1, 2019 | B.Braun Infusomat Space Volumetric Infusion Pump Administration Set, Catalog ... | Potential for the Infusion Pump Administration Set to leak and/or disconnect at the bonded joint ... | Class I | B. Braun Medical, Inc. |
| Apr 26, 2019 | SOLOPATH Balloon Expandable TransFemoral System, Model Numbers STFI-1425, S... | There is a potential for dislodgement of the tip from the outer diameter of the sheath resulting ... | Class I | Terumo Medical Corporation |
| Apr 26, 2019 | Portico Solo Re-Collapsible Access System , Model Numbers PRTSOLO-19, PRTSO... | There is a potential for dislodgement of the tip from the outer diameter of the sheath resulting ... | Class I | Terumo Medical Corporation |
| Apr 26, 2019 | SOLOPATH Re-Collapsible Access System, Model Numbers SR-1925, SR-1935, SR... | There is a potential for dislodgement of the tip from the outer diameter of the sheath resulting ... | Class I | Terumo Medical Corporation |
| Apr 11, 2019 | Straight Intraluminal Staplers, 25 mm diameter, Model SDH25A The Ethicon E... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Straight Intraluminal Staplers, 33 mm diameter, Model SDH33A The Ethicon E... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Endoscopic Curved Intraluminal Stapler, 21 mm diameter, Model ECS21A The E... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Endoscopic Curved Intraluminal Stapler, 33 mm diameter, Model ECS33A The E... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Straight Intraluminal Staplers, 29 mm diameter, Model SDH29A The Ethicon E... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Curved Intraluminal Staplers, 21 mm diameter, Model CDH21A The Ethicon End... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Endoscopic Curved Intraluminal Stapler, 25 mm diameter, Model ECS25A The E... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Straight Intraluminal Staplers, 21 mm diameter, Model SDH21A The Ethicon E... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Curved Intraluminal Staplers, 29 mm diameter, Model CDH29A The Ethicon End... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Curved Intraluminal Staplers, 25 mm diameter, Model CDH25A The Ethicon End... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Endoscopic Curved Intraluminal Stapler, 29 mm diameter, Model ECS29A The E... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 11, 2019 | Curved Intraluminal Staplers, 33 mm diameter, Model CDH33A The Ethicon End... | The staplers may have an insufficient firing stroke to break the washer and completely form staples. | Class I | Ethicon Endo-Surgery Inc |
| Apr 8, 2019 | MoniTorr INS1100 INS-1100 CSF Drainage System used with Pole Mount System ... | The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount sto... | Class I | Integra LifeSciences Corp. |
| Apr 8, 2019 | LimiTorr INS9020 INS-9020 LIMITORR VOLUME LIMITING EVD 20 ML LimiTorr INS902... | The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount sto... | Class I | Integra LifeSciences Corp. |
| Apr 8, 2019 | MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniT... | The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount sto... | Class I | Integra LifeSciences Corp. |
| Mar 28, 2019 | Edwards Lifesciences, Miller Balloon Atrioseptostomy Catheter, Sterile EO, Rx... | Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of ballo... | Class I | Edwards Lifesciences, LLC |
| Mar 28, 2019 | Edwards Lifesciences, Fogarty Dilation Atrioseptostomy Catheter, Sterile EO, ... | Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of ballo... | Class I | Edwards Lifesciences, LLC |
| Mar 25, 2019 | HAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTO... | New software version for affected ventilators reduces the probability of the ventilator entering ... | Class I | Hamilton Medical AG |