CareFusion Alaris Pump Module, Model 8100

FDA Device Recall #Z-1360-2020 — Class I — February 4, 2020

Recall Summary

Recall Number Z-1360-2020
Classification Class I — Serious risk
Date Initiated February 4, 2020
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm CareFusion 303, Inc.
Location San Diego, CA
Product Type Devices
Quantity 1286898

Product Description

CareFusion Alaris Pump Module, Model 8100

Reason for Recall

1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders

Distribution Pattern

Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ, MO, SD, UT, NM, MD, VT, MP, ND, AS; and countries of: Canada, Australia, New Zealand, the Middle East, South Africa, Taiwan, Southeast Asia and the European Union.

Lot / Code Information

Software versions 12.1.0, and 9.33 and prior

Other Recalls from CareFusion 303, Inc.

Recall # Classification Product Date
Z-1645-2026 Class II BD Alaris" System with Guardrails" Suite MX wit... Feb 11, 2026
Z-1028-2026 Class II BD Pyxis ES Enterprise Server Catalog Numbers 1... Dec 3, 2025
Z-0962-2026 Class II BD Pyxis MedStation ES, REF: 323, BD Pyxis M... Nov 20, 2025
Z-1217-2026 Class II BD Alaris" Pump Module model 8100 UDI-DI code:... Nov 6, 2025
Z-0430-2026 Class I BD Alaris infusion Pump Module 8100, REF: 8100;... Oct 17, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.