TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.

FDA Device Recall #Z-0164-2027 - Class II, September 16, 2026

Recall Summary

Recall Number Z-0164-2027
Classification Class II - Moderate risk
Date Initiated September 16, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Howmedica Osteonics Corp.
Location Mahwah, NJ
Product Type Devices
Quantity 605 trays

Product Description

TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.

Reason for Recall

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

Distribution Pattern

These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.

Lot / Code Information

Catalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted.

Other Recalls from Howmedica Osteonics Corp.

Recall # Classification Product Date
Z-0165-2027 Class II TS Femoral Prep Tray, Triathlon. Catalog Number... Sep 16, 2026
Z-0167-2027 Class II TS Misc. Instruments Tray, Triathlon. Catalog N... Sep 16, 2026
Z-0166-2027 Class II Triathlon Hinge Femoral Prep Tray. Catalog Numb... Sep 16, 2026
Z-3196-2026 Class II OMNIFIT SERIES I ACET. INSERT 10¿ 22mm ID Cat... Jul 30, 2026
Z-3206-2026 Class II SCORPIO C-DOME PATELLA 73-2710 73-2910 ... Jul 30, 2026

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately, you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.