SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-...
FDA Device Recall #Z-3206-2026 - Class II, July 30, 2026
Recall Summary
| Recall Number | Z-3206-2026 |
| Classification | Class II - Moderate risk |
| Date Initiated | July 30, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Howmedica Osteonics Corp. |
| Location | Mahwah, NJ |
| Product Type | Devices |
| Quantity | 12,156 US |
Product Description
SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason for Recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Lot / Code Information
73-2710 UDI-DI 07613327033847 Lot NPPV 73-2910 UDI-DI 07613327033854 Lot 14AGFD 73-3110 UDI-DI 07613327033861 Lots 1L45 4JM5 7EL5 AME3 MK74 NTV7 PYHM RE7J VEWJ 73-3508 UDI-DI 07613327033885 Lots HMM5 NLRE 73-3708 UDI-DI 07613327033892 Lots 3467 3D8D PMRHX1 T6YJ T9ME VWDA YAM7 73-3710 UDI-DI 07613327033908 Lots 3W25 DLYP 73-3910 UDI-DI 07613327033915 Lots 14NEYA 1XED 9Y3T D1M9 D2YP EX8T KRT3 KRT3X1 T325 V8X8 X6YE
Other Recalls from Howmedica Osteonics Corp.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0166-2027 | Class II | Triathlon Hinge Femoral Prep Tray. Catalog Numb... | Sep 16, 2026 |
| Z-0165-2027 | Class II | TS Femoral Prep Tray, Triathlon. Catalog Number... | Sep 16, 2026 |
| Z-0164-2027 | Class II | TS Universal Baseplate Finishing Tray, Triathlo... | Sep 16, 2026 |
| Z-0167-2027 | Class II | TS Misc. Instruments Tray, Triathlon. Catalog N... | Sep 16, 2026 |
| Z-3204-2026 | Class II | TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D ... | Jul 30, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.