LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-0012...
FDA Device Recall #Z-2065-2026 — Class II — April 13, 2026
Recall Summary
| Recall Number | Z-2065-2026 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 13, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Physio-Control, Inc. |
| Location | Redmond, WA |
| Product Type | Devices |
| Quantity | 206 |
Product Description
LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122
Reason for Recall
Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.
Distribution Pattern
US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.
Lot / Code Information
Device/UDI/Serial Number: LIFEPAK 15/0088387391164815011543166573, 0088387387169015033143394891, 0088387387169015050743524360, 0088387387169015092843914574, 0088387387204815093043915122, 0088387387169016010444187447, 0088387387162116012044252611, 0088387387169016012644262721, 0088387387169016012644277082, 0088387387169016012744282280, 0088387387171316060244574340, 0088387387171316080444896329, 0088387387169016083144962804, 0088387387170617013145447229, 0088387387169017041245743786, 0088387387171317050945834209, 0088387387171317062845998142, 0088387387171317082846103043, 0088387387169017101046303288, 0088387387171317103146303092, 0088387387162117122146587130, 0088387387171317122646602003, 0088387387171317122846608783, 0088387387171317122946614788, 0088387387171318022246813507, 0088387387169018032946966267, 0088387387171318061947337396, 0088387387169018070547401928, 0088387387171318092747611441, 0088387387169018112947784931, 0088387387169018112947786317, 0088387387171318121847847661, 0088387387171318121847848950, 0088387387708118121947845317, 0088387387708118121947855371, 0088387387169018122047857172, 0088387387169018122047858660, 0088387387171318122847872868, 0088387387171319031248018370, 0088387387169019061248247559, 0088387387168319062748298704, 0088387387169019100848453060, 0088387387171319101448455890, 0088387387171319121848586351, 0088387387171319121948586599, 0088387387168320012848651507, 0088387387171320030548701689, 0088387387708120042848784975, 0088387387171320071648887706, 0088387387168320081748921846, 0088387387169020092448981398, 0088387387169020102649032288, 0088387387169020102649050140, 0088387387162120110249055931, 0088387387168320122949156402, 0088387387169021011149162067, 0088387387169021011149167463, 0088387387169021013049212578, 0088387387169021020949225869, 0088387387169021020949226060, 0088387387176821021949240566, 0088387387168321033149345530, 0088387387169021042149378981, 0088387387169021042649381781, 0088387387169021042749381576, 0088387387168321081849559854, 0088387387168321081949560127, 0088387387164521091649607224, 0088387387171321102149662806, 0088387377697122070750045009, 0088387387169022090150136719, 0088387387171322090850152289, 0088387387708122091350158293, 0088387387169022101950223934, 0088387387708122112150288760, 0088387387171322120150307353, 0088387387169023062850629377, 0088387387169023072450653772, 0088387387169023072450667197, 0088387387708123072850669746, 0088387387708123082850714419, 0088387387171323082950720588, 0088387387169023112150850129, 0088387387708123122050883524, 0088387388434824052251092313, 0088387387170624121151422867, 0088387387170624122351440213, 0088387387170624122351444039, 0088387388434824052251092313, 0088387387169019100848453060, 0088387387201715090843839264, 0088387387171318121847850793/ 38523489, 39736196, 40465345, 40717931, 40765749, 41130420, 41139184, 41420681, 41446039, 41446693, 41704583, 42221131, 42920956, 42921954, 43166573, 43394891, 43524360, 43914574, 43915122, 44187447, 44252611, 44262721, 44277082, 44282280, 44574340, 44896329, 44962804, 45447229, 45743786, 45834209, 45998142, 46103043, 46303288, 46303092, 46587130, 46602003, 46608783, 46614788, 46813507, 46966267, 47337396, 47401928, 47611441, 47784931, 47786317, 47847661, 47848950, 47845317, 47855371, 47857172, 47858660, 47872868, 48018370, 48247559, 48298704, 48453060, 48455890, 48586351, 48586599, 48651507, 48701689, 48784975, 48887706, 48921846, 48981398, 49032288, 49050140, 49055931, 49156402, 49162067, 49167463, 49212578, 49225869, 49226060, 49240566, 49345530, 49378981, 49381781, 49381576, 49559854, 49560127, 49607224, 49662806, 50045009, 50136719, 50152289, 50158293, 50223934, 50288760, 50307353, 50629377, 50653772, 50667197, 50669746, 50714419, 50720588, 50850129, 50883524, 51092313, 51422867, 51440213, 51444039, 51092313, 48453060, 43839264, 47850793, 38369584. LIFEPAK 1000/0088507469887815020943235807, 0088507469895315031343346860, 0088507469887815111944068667, 0088507469887815111944068669, 0088507469895316011844239775, 0088507469887816041444539726, 0088507469887817061745966118, 0088507469887817061745966121, 0088507469887817061745966122, 0088507469887817061745966126, 0088507469887817061945966131, 0088507469887817062945994506, 0088507469887817062946004311, 0088507469887817063046012479, 0088507469887818060547291148, 0088387381546519080648366102, 0088387381545820092948987376, 0088387381545821011249174469, 0088387381545821081349544565, 0088387381545821090249586700, 0088387381546521122249752235, 0088387381545822012649794992, 0088387381545822032949882696, 0088387381545822032949882698, 0088387381545822032949882819/ 34933095, 35096556, 35184465, 36153943, 36199099, 36560529, 36560531, 37119097, 37226402, 37330295, 38074718, 38074722, 38074727, 38074737, 38074741, 38074744, 38074745, 38074747, 38074748, 38074749, 38077938, 38077939, 38077940, 38077941, 38077944, 38378802, 38513659, 38656117, 38656121, 38656122, 38957375, 38957392, 38957397, 38957399, 38957402, 38972629, 38972677, 39775405, 39775406, 39775407, 39775415, 39964908, 40323073, 40412729, 40541187, 40584085, 40594714, 40707896, 40725568, 40725569, 40725571, 40725572, 40725575, 40725584, 40725582, 40725583, 41306695, 41692754, 41863786, 42504017, 42520778, 42803174, 42921847, 43235807, 43346860, 44068667, 44068669, 44239775, 44539726, 45966118, 45966121, 45966122, 45966126, 45966131, 45994506, 46004311, 46012479, 47291148, 48366102, 48987376, 49174469, 49544565, 49586700, 49752235, 49794992, 49882696, 49882698, 49882819. LIFEPAK 35/0088387377265224042251041190, 0088387377265224041851042010, 0088387377265224042251044581, 0088387376560924071151156570, 0088387376560924071151157072, 0088387376560924071151157449, 0088387376560924080651182632, 0088387376560924091251263526, 0088387376560924121051417585, 0088387376560925020451496956/ 51041190, 51042010, 51044581, 51156570, 51157072, 51157449, 51182632, 51263526, 51417585, 51496956. LIFEPAK 20e/0088387385386320082448935344, 0088387385386320082448935408, 0088387385386321080249537075/ 48935344, 48935408, 49537075
Other Recalls from Physio-Control, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1544-2026 | Class II | PHYSIO CONTROL INFANT/CHILD Reduced Energy Defi... | Jan 21, 2026 |
| Z-2498-2025 | Class II | LIFEPAK 1000 Part / Catalog Numbers / UDI-DI co... | Jul 22, 2025 |
| Z-2499-2025 | Class II | LIFEPAK CR2 Part / Catalog Numbers / UDI-DI co... | Jul 22, 2025 |
| Z-2497-2025 | Class II | LIFEPAK 15 Part / Catalog Numbers / UDI-DI code... | Jul 22, 2025 |
| Z-1232-2025 | Class II | LIFEPAK 20e REF 99507-000001 The AED mode is... | Feb 3, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.