PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-00001...
FDA Recall #Z-1544-2026 — Class II — January 21, 2026
Product Description
PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).
Reason for Recall
Due to pediatric defibrillator electrode delamination
Recalling Firm
Physio-Control, Inc. — Redmond, WA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
16,499 electrodes (US/OUS: 8327/8172)
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Australia, Canada, Hong Kong, Japan, Mexico, Malaysia, Netherlands, New Zealand, and Taiwan.
Code Information
Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Lot Numbers: 315303, 315639, 315858, 316007, 320529, 318931
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated