Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper ...
FDA Recall #Z-2522-2025 — Class II — August 6, 2025
Product Description
Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper included. 50mL, Product Code 21-7001-24
Reason for Recall
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
Recalling Firm
ICU Medical Inc. — Minneapolis, MN
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
29811 units
Distribution
Worldwide distribution.
Code Information
UDI/DI 10610586022111, Lot Numbers: 6022008, 6022009, 6037749, 6062686, 6062687
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated