Smiths Medical NRFit CADD Yellow Medication Cassette Reservoir with NRFit connector with flow sto...
FDA Recall #Z-2527-2025 — Class II — August 6, 2025
Product Description
Smiths Medical NRFit CADD Yellow Medication Cassette Reservoir with NRFit connector with flow stop, yellow-striped tubing, clamp and female NRFit connector. Yellow stopper included 100mL", Product Code 21-7600-24
Reason for Recall
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
Recalling Firm
ICU Medical Inc. — Minneapolis, MN
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
15546 units
Distribution
Worldwide distribution.
Code Information
UDI/DI 10610586044007, Lot Numbers: 6037784, 6054015
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated