daVinci X EndoWrist(R) Stapler 45, REF 470298 Product Usage: The Intuitive Surgical EndoWrist...

FDA Device Recall #Z-0404-2018 — Class II — September 19, 2017

Recall Summary

Recall Number Z-0404-2018
Classification Class II — Moderate risk
Date Initiated September 19, 2017
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Intuitive Surgical, Inc.
Location Sunnyvale, CA
Product Type Devices
Quantity 60 units

Product Description

daVinci X EndoWrist(R) Stapler 45, REF 470298 Product Usage: The Intuitive Surgical EndoWrist Stapler 45, Stapler 45 Reloads, and other Stapler Accessories (including the bladeless obturators) are intended to be used with the da Vinci Surgical System (Model IS4000) for resection, transection, and/or creation of anastomoses in General, Thoracic, Gynecologic and Urologic surgery. The device can be used with the same staple line or tissue buttressing material (natural synthetic).

Reason for Recall

Specific da Vinci Xi EndoWrist Stapler 45 instruments are potentially impacted by a variation in the manufacturing process. In affected instruments, there is a possibility that the Stapler Release Kit (SRK) Wrench interface feature accessed through release hole 2 is manufactured to an incorrect dimension, which may prevent the SRK Wrench from being able to manually unclamp the instrument.

Distribution Pattern

US Nationwide Distribution

Lot / Code Information

UDI 00886874112427: Serial Numbers: S10170630, S11170705, S10170707, S10170731

Other Recalls from Intuitive Surgical, Inc.

Recall # Classification Product Date
Z-1355-2026 Class II da Vinci 5 Surgeon Console Viewer Display, ASSY... Dec 30, 2025
Z-1096-2026 Class II da Vinci SP, SP Access Port Kit, Large Incisio... Dec 15, 2025
Z-1134-2026 Class II Brand Name: Ion Endoluminal System Product Nam... Dec 4, 2025
Z-1428-2026 Class II Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/... Nov 17, 2025
Z-1232-2026 Class II Brand Name: Da Vinci Product Name: da Vinci X,... Nov 7, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.