da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

FDA Recall #Z-1355-2026 — Class II — December 30, 2025

Recall #Z-1355-2026 Date: December 30, 2025 Classification: Class II Status: Ongoing

Product Description

da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

Reason for Recall

Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

Recalling Firm

Intuitive Surgical, Inc. — Sunnyvale, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

47

Distribution

US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA.

Code Information

UDI: 00886874119747, Serial: SQ0326, SQ0336, SQ0341, SQ0345, SQ0346, SQ0356, SQ0367, SQ0368, SQ0375, SQ0377, SQ0379, SQ0380, SQ0386, SQ0387, SQ0389, SQ0394, SQ0395, SQ0398, SQ0401, SQ0404, SQ0412, SQ0413, SQ0415, SQ0419, SQ0420, SQ0422, SQ0425, SQ0427, SQ0428, SQ0432, SQ0433, SQ0436, SQ0443, SQ0447, SQ0448, SQ0451, SQ0454, SQ0457, SQ0458, SQ0462, SQ0464, SQ0469, SQ0470, SQ0662, SQ0727, SQ0740, SQ0761

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated