Fabius Tiro M, anesthesia machine, catalog no. 8608400 Product Usage: The products are inhala...

FDA Device Recall #Z-1484-2018 — Class I — March 1, 2018

Recall Summary

Recall Number Z-1484-2018
Classification Class I — Serious risk
Date Initiated March 1, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Draeger Medical, Inc.
Location Telford, PA
Product Type Devices
Quantity 1

Product Description

Fabius Tiro M, anesthesia machine, catalog no. 8608400 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).

Reason for Recall

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

Distribution Pattern

US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.

Lot / Code Information

Serial no. ASKK-0001

Other Recalls from Draeger Medical, Inc.

Recall # Classification Product Date
Z-0668-2025 Class I Brand Name: Atlan Product Name: Atlan A300XL ... Oct 15, 2024
Z-0304-2025 Class I Brand Name: Atlan Product Name: Atlan A350 Mo... Oct 15, 2024
Z-0305-2025 Class I Brand Name: Atlan Product Name: Atlan A350XL ... Oct 15, 2024
Z-0667-2025 Class I Brand Name: Atlan Product Name: Atlan A300 ... Oct 15, 2024
Z-1512-2024 Class I Dr¿ger Perseus A500 Anesthesia Workstation Mar 20, 2024

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.