VirtuoSaph Plus Endoscopic Vessel Harvesting System, Catalog Number: VSP550EX Product Usage: ...
FDA Device Recall #Z-0227-2019 — Class II — August 31, 2018
Recall Summary
| Recall Number | Z-0227-2019 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 31, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Terumo Cardiovascular Systems Corporation |
| Location | Elkton, MD |
| Product Type | Devices |
| Quantity | 135 |
Product Description
VirtuoSaph Plus Endoscopic Vessel Harvesting System, Catalog Number: VSP550EX Product Usage: The VittuoSaph¿ Plus Endoscopic Vessel Harvesting System VSP550EX is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surge1y for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requit¿ing blunt dissection of tissue including dissection of blood vessels and dissection of blood vessels of the extremities. Extremity procedures include tissue dissection and/or vessel harvesting along the saphenous vein for coronary artery bypass grafting and peripheral artery bypass grafting or radial artery for use in coronary artery bypass grafting vc
Reason for Recall
The system's IFU contains a list of generators that the have been deemed compatible. Recent reports indicate that one energy platform is not compatible with the system. Therefore, it has been removed from the list of compatible generators within the IFU.
Distribution Pattern
Worldwide - US Nationwide in the states of FL, IL, NY, OH, TX. The products were distributed to the following foreign country: Singapore.
Lot / Code Information
Lot Numbers: 81K
Other Recalls from Terumo Cardiovascular Systems Corpora...
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2018-2025 | Class II | CDI OneView System BPM Probe: Cat. No. CDI75... | May 28, 2025 |
| Z-1492-2025 | Class II | Terumo CDI OneView Monitoring System Hematocrit... | Mar 3, 2025 |
| Z-1401-2024 | Class II | Cardiovascular Procedure Kits that include High... | Feb 20, 2024 |
| Z-1346-2023 | Class II | CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Nu... | Mar 1, 2023 |
| Z-1347-2023 | Class II | CAPIOX FX25 Hollow Fiber Oxygenator, Catalog Nu... | Mar 1, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.