CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CD...

FDA Recall #Z-2018-2025 — Class II — May 28, 2025

Recall #Z-2018-2025 Date: May 28, 2025 Classification: Class II Status: Ongoing

Product Description

CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit

Reason for Recall

Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.

Recalling Firm

Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

28 US; 7 OUS

Distribution

US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.

Code Information

UDI-DI 00886799001889 Lot Number/Serial Number Range: B0050231-B0050240 B0050221-B0050230 B0050211-B0050219 B0050201-B0050209 B0050191-B0050200

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated