CURVED-TIP STAPLER 30,IS4000, model number 470530; Gastroenterology/Urology: The Intuitive Su...
FDA Device Recall #Z-2350-2016 — Class II — July 14, 2016
Recall Summary
| Recall Number | Z-2350-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | July 14, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Intuitive Surgical, Inc. |
| Location | Sunnyvale, CA |
| Product Type | Devices |
| Quantity | 2065 total |
Product Description
CURVED-TIP STAPLER 30,IS4000, model number 470530; Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device is indicated for adult and pediatric use. The device can be used with staple line or tissue buttressing material.
Reason for Recall
An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend
Distribution Pattern
US, including Puerto Rico and Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Monaco, Netherlands, Norway, Portugal, , Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Venezuela.
Lot / Code Information
All
Other Recalls from Intuitive Surgical, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1355-2026 | Class II | da Vinci 5 Surgeon Console Viewer Display, ASSY... | Dec 30, 2025 |
| Z-1096-2026 | Class II | da Vinci SP, SP Access Port Kit, Large Incisio... | Dec 15, 2025 |
| Z-1134-2026 | Class II | Brand Name: Ion Endoluminal System Product Nam... | Dec 4, 2025 |
| Z-1428-2026 | Class II | Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/... | Nov 17, 2025 |
| Z-1232-2026 | Class II | Brand Name: Da Vinci Product Name: da Vinci X,... | Nov 7, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.