VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Cli...

FDA Device Recall #Z-1765-2017 — Class II — March 7, 2017

Recall Summary

Recall Number Z-1765-2017
Classification Class II — Moderate risk
Date Initiated March 7, 2017
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm ORTHO-CLINICAL DIAGNOSTICS
Location Bridgend, N/A
Product Type Devices
Quantity US: 6324; Foreign: 1868

Product Description

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK

Reason for Recall

Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.

Distribution Pattern

Worldwide: United States, Puerto Rico, Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, and Venezuela.

Lot / Code Information

Lot 1570 (expiry 22 March 2017), Lot 1580 (expiry 13 April 2017), Lot 1590 (expiry 27 April 2017)

Other Recalls from ORTHO-CLINICAL DIAGNOSTICS

Recall # Classification Product Date
Z-1372-2024 Class II VITROS Performance Verifier I Feb 27, 2024
Z-1373-2024 Class II VITROS Performance Verifier II Feb 27, 2024
Z-2358-2023 Class II VITROS Immunodiagnostic Products Troponin I ES ... Jun 5, 2023
Z-2357-2023 Class II VITROS Immunodiagnostic Products Troponin I ES ... Jun 5, 2023
Z-2129-2023 Class II VITROS Chemistry Products Calibrator Kit 20-in ... May 8, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.