VITROS Performance Verifier I
FDA Recall #Z-1372-2024 — Class II — February 27, 2024
Product Description
VITROS Performance Verifier I
Reason for Recall
The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
Recalling Firm
Ortho-Clinical Diagnostics — Rochester, NY
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
10,911 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Code Information
Product Code: 8067324; GTIN: 10758750004317; Lot No. Q1174.
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated