Turon Shoulder Impaction Fixture REF 804-15-102 The Turon and RSP Impaction Fixtures are used in...
FDA Device Recall #Z-0179-2016 — Class II — September 14, 2015
Recall Summary
| Recall Number | Z-0179-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | September 14, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Encore Medical, Lp |
| Location | Austin, TX |
| Product Type | Devices |
| Quantity | 277 units |
Product Description
Turon Shoulder Impaction Fixture REF 804-15-102 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.
Reason for Recall
During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
Distribution Pattern
Nationwide Distribution
Lot / Code Information
Lot # 52748L02A
Other Recalls from Encore Medical, Lp
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1459-2026 | Class II | Brand Name: EMPOWR 3D KNEE Product Name: EMPOW... | Jan 15, 2026 |
| Z-1458-2026 | Class II | Brand Name: Reverse Shoulder Prosthesis (RSP) ... | Jan 15, 2026 |
| Z-1460-2026 | Class II | Brand Name: Reverse Shoulder Prosthesis (RSP) ... | Jan 15, 2026 |
| Z-1234-2026 | Class II | EMPOWR Acetabular System: Liner, 10 degree Hood... | Jan 5, 2026 |
| Z-1233-2026 | Class II | EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: ... | Jan 5, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.