EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree H...

FDA Recall #Z-1234-2026 — Class II — January 5, 2026

Recall #Z-1234-2026 Date: January 5, 2026 Classification: Class II Status: Ongoing

Product Description

EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded, HXe+, 40G, REF: 942-01-40G; Liner, 10 degree Hooded, HXe+, 32C, REF: 942-01-32C

Reason for Recall

942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.

Recalling Firm

Encore Medical, LP — Austin, TX

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

49

Distribution

US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CA

Code Information

REF/UDI-DI/Lot(Expiration): 942-01-36H/00190446309789/624Z1032(03-15-2026), 942-01-36J/00190446309802/626Z1021(03-15-2026), 942-01-40G/00190446309826/628Z1034(03-15-2026), 942-01-32C/00190446309697/615Z1026(03-15-2026)

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated