0 deg, 8mm Endoscope, IS4000: Model numbers: 470026-34, 470026-40, 470026-41. General and Pla...

FDA Device Recall #Z-0652-2016 — Class II — November 23, 2015

Recall Summary

Recall Number Z-0652-2016
Classification Class II — Moderate risk
Date Initiated November 23, 2015
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Intuitive Surgical, Inc.
Location Sunnyvale, CA
Product Type Devices
Quantity 2624 total all models

Product Description

0 deg, 8mm Endoscope, IS4000: Model numbers: 470026-34, 470026-40, 470026-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on the Surgeon Console 3D viewer and Vision Cart touchscreen. Endoscopes are available with either a 0 degree or 30 degree tips.

Reason for Recall

This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm Endoscopes in combination with the da Vinci 8 mm and 12 mm cannulae, cannula reducers, and monopolar energy instruments. It informs customers of scenarios that may result in unintended patient harm due to capacitive coupling from monopolar energy sources.

Distribution Pattern

Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela. .

Lot / Code Information

Model numbers: 470026-34, 470026-40, 470026-41. All lots.

Other Recalls from Intuitive Surgical, Inc.

Recall # Classification Product Date
Z-1355-2026 Class II da Vinci 5 Surgeon Console Viewer Display, ASSY... Dec 30, 2025
Z-1096-2026 Class II da Vinci SP, SP Access Port Kit, Large Incisio... Dec 15, 2025
Z-1134-2026 Class II Brand Name: Ion Endoluminal System Product Nam... Dec 4, 2025
Z-1428-2026 Class II Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/... Nov 17, 2025
Z-1232-2026 Class II Brand Name: Da Vinci Product Name: da Vinci X,... Nov 7, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.