Habib EUS RFA, 10mm electrode active tip, Material Number/UPN: 6500 - Product Usage: The Habib EU...

FDA Device Recall #Z-0667-2020 — Class II — November 18, 2019

Recall Summary

Recall Number Z-0667-2020
Classification Class II — Moderate risk
Date Initiated November 18, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Boston Scientific Corporation
Location Marlborough, MA
Product Type Devices
Quantity 139

Product Description

Habib EUS RFA, 10mm electrode active tip, Material Number/UPN: 6500 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.

Reason for Recall

During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, CT, FL, NJ, NY, MA, MD, MN, TN, TX and countries of Canada, India, Australia, Great Britain, Greece, Spain.

Lot / Code Information

Material Number/UPN: 6500 GTIN: 5060271140725 Batches: 17-C02 (Expiration: 9-Jul-2020), 17-C06 (Expiration: 19-Feb-2021), 17-C07 (Expiration: 19-Feb-2021), 17-C08 (Expiration: 20-Feb-2021)

Other Recalls from Boston Scientific Corporation

Recall # Classification Product Date
Z-1551-2026 Class II EndoVive Safety PEG Kit; Outer Box Material Nu... Feb 10, 2026
Z-1160-2026 Class I AXIOS Stent and Electrocautery-Enhanced Deliver... Dec 19, 2025
Z-1162-2026 Class I *HOT AXIOS Stent and Electrocautery-Enhanced De... Dec 19, 2025
Z-1159-2026 Class I AXIOS Stent and Electrocautery-Enhanced Deliver... Dec 19, 2025
Z-1163-2026 Class I *HOT AXIOS Stent and Electrocautery-Enhanced De... Dec 19, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.