EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number ...

FDA Recall #Z-1551-2026 — Class II — February 10, 2026

Recall #Z-1551-2026 Date: February 10, 2026 Classification: Class II Status: Ongoing

Product Description

EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;

Reason for Recall

Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit.

Recalling Firm

Boston Scientific Corporation — Marlborough, MA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

806 units

Distribution

US Nationwide distribution.

Code Information

Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; Outer Box UDI: 08714729880394; Inner Box UDI: 08714729880387; Outer Box Lot#s: 37435463, 37522541, 37528545, 37546847, 37546850, 37546849, 37550500, 37592036, 37732292; Inner Box Lot#s: 37435462, 37514809, 37525231, 37536563, 37536564, 37536562, 37546840, 37583642, 37731897;

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated