da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 0088687411...

FDA Device Recall #Z-0283-2021 — Class II — June 17, 2019

Recall Summary

Recall Number Z-0283-2021
Classification Class II — Moderate risk
Date Initiated June 17, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Intuitive Surgical, Inc.
Location Sunnyvale, CA
Product Type Devices
Quantity 1 system

Product Description

da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number: 381121-31 (UDI 00886874110898) - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.

Reason for Recall

One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images.

Distribution Pattern

US Nationwide distribution including in the state of Washington.

Lot / Code Information

IS4000 Endoscope Controller, Serial Number: 432242 installed on the da Vinci Surgical System, Serial Number: SK2137

Other Recalls from Intuitive Surgical, Inc.

Recall # Classification Product Date
Z-1355-2026 Class II da Vinci 5 Surgeon Console Viewer Display, ASSY... Dec 30, 2025
Z-1096-2026 Class II da Vinci SP, SP Access Port Kit, Large Incisio... Dec 15, 2025
Z-1134-2026 Class II Brand Name: Ion Endoluminal System Product Nam... Dec 4, 2025
Z-1428-2026 Class II Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/... Nov 17, 2025
Z-1232-2026 Class II Brand Name: Da Vinci Product Name: da Vinci X,... Nov 7, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.